Embolization of intracranial aneurysms with hydrogel-coated coils versus inert platinum coils: effects on packing density, coil length and quantity, procedure performance, cost, length of hospital stay, and durability of therapy.

نویسندگان

  • Ron C Gaba
  • Sameer A Ansari
  • Soma Sinha Roy
  • Franklin A Marden
  • Marlos A G Viana
  • Tim W Malisch
چکیده

BACKGROUND AND PURPOSE The durability of aneurysm coil embolization is thought to depend on packing density. The expansile property of hydrogel coating on coils increases volumetric packing per coil length. We describe our experience using hydrogel-coated coils (HydroCoils) compared with inert platinum coils in intracranial aneurysm embolization. METHODS Fifty aneurysms embolized primarily using HydroCoils from 2003 to 2004 were compared with 57 volume- and shape-matched aneurysms treated with standard platinum coils from 2000 to 2003. Outcome measures included volumetric percentage occlusion (VPO), length and number of coils used, procedure time, fluoroscopy time, contrast volume, coil cost, length of hospital stay, and durability of therapy. RESULTS Seventeen/26/5 small/medium/large aneurysms treated with HydroCoils were matched with 29/24/4 small/medium/large aneurysms treated with inert platinum. HydroCoil embolization yielded significantly greater VPO (84.8% versus 29.8%; P<0.001), decreased average total coil length used per aneurysm (33.2 versus 44.3 cm), reduced fluoroscopy time (53.2 versus 65.2 minutes; P=0.016), but increased contrast volume used (174.8 versus 112.9 cc; P<0.001). There were no differences in length of hospital stay. Procedure-related morbidity and mortality rates in the HydroCoil cohort were 4% and 0%, respectively. Follow-up angiography at mean 12.3 months revealed lower aneurysm recurrence rates (17% versus 24%; number-needed-to-treat [NNT] 14.3). Initial costs associated with HydroCoil embolization were higher ($5835 versus $4017; P=0.004) but countered by lower retreatment rates (10% versus 17%; NNT 14.3). CONCLUSIONS HydroCoil embolization achieves greater aneurysm packing density with decreased coil length. Initial durability data favor HydroCoils, with lower recurrence and retreatment rates.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Endovascular Embolization of Intracranial Aneurysms Using Bare Platinum Axium™ Detachable Coils: Immediate and Short-Term Follow-up Results from a Multicenter Registry

PURPOSE Axium™ coils were developed to improve the durability of coil-embolized cerebral aneurysms by increasing packing density. The purpose of this prospective multicenter registry was to evaluate the safety and durability of Axium™ coils. MATERIALS AND METHODS One hundred twenty-six patients with 135 aneurysms of ≤ 15 mm in size underwent coil embolization using bare platinum coils, with A...

متن کامل

Treatment of unruptured intracranial aneurysms using internally expanding coils.

BACKGROUND AND PURPOSE The International Subarachnoid Aneurysm Trial (ISAT) showed that patients with intracranial aneurysms treated with coil embolization have better clinical outcomes than those undergoing neurosurgical clipping. However some patients treated endovascularly have recurrence of aneurysms. Low packing density is often cited as a reason for recurrence. Coiling with hydrogel cover...

متن کامل

Matrix² coils in embolization of intracranial aneurysms: 1-year outcome and comparison with bare platinum coil group in a single institution.

BACKGROUND AND PURPOSE The endosaccular occlusion by using BPC has been useful in the treatment of intracranial aneurysms, but its limited durability remains a deep-seated drawback. The Matrix² coil, one of the bioactive-coated coils, had been developed to improve this limited durability. To evaluate durability of Matrix² coils after embolization of intracranial aneurysms, we retrospectively co...

متن کامل

Short-term outcome of intracranial aneurysms treated with polyglycolic acid/lactide copolymer-coated coils compared to historical controls treated with bare platinum coils: a single-center experience.

BACKGROUND AND PURPOSE Aneurysm recanalization is an innate problem in endovascular treatment of aneurysms with coils. A coated coil system, covered with a bioabsorbable polymeric material (polyglycolic acid/lactide copolymer, PGLA), was developed to accelerate intra-aneurysmal clot organization and fibrosis. The purpose of this study was to evaluate the efficacy and safety of the PGLA-coated c...

متن کامل

Angiographic and histologic analysis of experimental aneurysms embolized with platinum coils, Matrix, and HydroCoil.

BACKGROUND AND PURPOSE A report directly comparing platinum coils, Matrix coils, and HydroCoils in a single animal model does not currently exist. We evaluated and compared the performance of these three products in the embolization of experimental aneurysms. METHODS Thirty-three elastase-induced saccular aneurysms were created in rabbits. Aneurysms were embolized with Matrix coils (n = 15), ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Stroke

دوره 37 6  شماره 

صفحات  -

تاریخ انتشار 2006