3D echocardiographic location of implantable device leads and mechanism of associated tricuspid regurgitation.
نویسندگان
چکیده
OBJECTIVES This study sought to: 1) determine the feasibility of using 3-dimensional transthoracic echocardiography (3D TTE) in patients with implantable cardiac resynchronization devices, pacemakers, and defibrillators to visualize the device leads in the right heart and their position relative to the tricuspid valve leaflets; 2) determine the prevalence of different lead positions; and 3) study the relationship between lead location and tricuspid regurgitation (TR) severity. BACKGROUND Pacemaker, defibrillator, and cardiac resynchronization device implantation is currently guided by fluoroscopy, not allowing targeted lead positioning relative to the tricuspid valve leaflets. These leads have been reported to cause TR of variable degrees, but echocardiography is not routinely used to elucidate the mechanisms of lead interference with tricuspid valve leaflets in individual patients. METHODS 3D TTE full-volume images of the right ventricle and/or zoomed images of the tricuspid valve were obtained in 121 patients with implanted devices. Images were viewed offline to determine the position of the device-lead relative to the tricuspid valve leaflets. Severity of TR was estimated on the basis of vena contracta measurements. RESULTS 3D TTE clearly depicted lead position in 90% of patients. The right ventricular lead was impinging on either the posterior (20%) or septal (23%) leaflet or was not interfering with leaflet motion (53%) when positioned near the posteroseptal commissure or in the central portion of the tricuspid valve orifice. In the remaining patients, leads were impinging on the anterior leaflet (4%) or positioned in either the anteroposterior or anteroseptal commissure (3%). Leads interfering with normal leaflet mobility were associated with more TR than nonimpinging leads (vena contracta: median 0.62 cm [1st and 3rd quartiles: 0.51, 0.84 cm] vs. 0.27 cm [1st and 3rd quartiles: 0.00, 0.48 cm]; p < 0.001). CONCLUSIONS 3D TTE showed a clear association between device lead position and TR. To minimize TR induced by device-leads, 3D TTE guidance should be considered for placement in a commissural position.
منابع مشابه
Impact of implantable transvenous device lead location on severity of tricuspid regurgitation.
BACKGROUND Implantable device leads can cause tricuspid regurgitation (TR) when they interfere with leaflet motion. The aim of this study was to determine whether lead-leaflet interference is associated with TR severity, independent of other causative factors of functional TR. METHODS A total of 100 patients who underwent transthoracic two-dimensional and three-dimensional (3D) echocardiograp...
متن کاملRisk factors for mortality in patients with cardiac device-related infection.
BACKGROUND Because of the increased use of pacemakers and implantable cardioverter defibrillators, infection has become a complication with significant morbidity and mortality. Data on risk factors for mortality in patients with cardiac-device related infection are limited. We evaluated the prognostic significance of key clinical and echocardiographic variables in a large retrospective populati...
متن کاملSevere symptomatic tricuspid valve regurgitation due to permanent pacemaker or implantable cardioverter-defibrillator leads.
OBJECTIVES We report a series of patients with severe tricuspid valve regurgitation due to a permanent pacemaker (PPM) or implantable cardioverter-defibrillator (ICD) lead. BACKGROUND Severe tricuspid regurgitation caused by a PPM or ICD lead is an under-recognized but treatable etiology of severe right heart failure. METHODS We reviewed the records of 41 patients who underwent tricuspid va...
متن کاملDetection of tricuspid regurgitation with two-dimensional echocardiography and peripheral vein injections.
"Contrast echocardiography", utilizing a two-dimensional ultrasound system and peripheral venous injections, was used for the detection of tricuspid regurgitation in 30 patients. The appearance of contrast in the inferior vena cava and the back and forth movement of contrast across the tricuspid valve were considered evidence for tricuspid regurgitation. Echocardiographic findings were correlat...
متن کامل3D imaging of device leads: "taking the lead with 3D".
Pacemaker lead–associated tricuspid regurgitation (TR) has been anecdotally described for many years, but investigation has been limited to a small number of observational studies (1). Overall, it seems that worsening of TR is common after transvenous lead placement (2). Perhaps the lack of formal research in this area has stemmed from our previous inability to adequately confirm that device le...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- JACC. Cardiovascular imaging
دوره 7 4 شماره
صفحات -
تاریخ انتشار 2014