Essential opaque bubble layer elimination with novel LASIK flap settings in the FS200 Femtosecond Laser

نویسندگان

  • A John Kanellopoulos
  • George Asimellis
چکیده

BACKGROUND The purpose of this study is to evaluate the extent and incidence of opaque bubble layer (OBL) using laser-assisted in situ keratomileusis (LASIK) flaps created with the Alcon/WaveLight® FS200 femtosecond laser as a result of a recent change in flap programming parameters aiming to reduce further the incidence and extent of OBL. METHODS Intraoperative digital images of flaps from 36 consecutive patients (72 eyes) subjected to bilateral femtosecond-assisted LASIK were analyzed using a proprietary computerized technique. The incidence and extent of OBL was measured and reported as a percentage of the entire flap area. Flap creation was performed with a 1.7 mm wide canal, implemented as an updated design intended to reduce the extent of OBL (group A). The same OBL parameters were investigated and compared in an age-matched and procedure-matched patients in whom the previous standard setting of a 1.3 mm wide canal was implemented (group B). RESULTS In group A, the average extent of OBL was 3.69% of the flap area (range 0%-11.34%). In group B, the respective values were 6.06% (range 0%-20.24%). We found the difference to be statistically significant (one-tailed P = 0.00452). CONCLUSION This study suggests that there is a significant reduction in the incidence and extent of OBL when novel LASIK flap ventilation canal parameters of width and spot line separation are used.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique

BACKGROUND The purpose of this study was to determine flap parameter accuracy, extent of the opaque bubble layer, and incidence of skip lines in femtosecond laser-assisted stromal in situ keratomileusis (LASIK) using the WaveLight(®) FS200 laser and optoelectronic clinical measurements. METHODS Images from 101 flaps were automatically recorded during consecutive routine LASIK procedures perfo...

متن کامل

High incidence of rainbow glare after femtosecond laser assisted-LASIK using the upgraded FS200 femtosecond laser

BACKGROUND To compare the incidence of rainbow glare (RG) after femtosecond laser assisted-LASIK (FS-LASIK) using the upgraded FS200 femtosecond laser with different flap cut parameter settings. METHODS A consecutive series of 129 patients (255 eyes) who underwent FS-LASIK for correcting myopia and/or astigmatism using upgraded WaveLight FS200 femtosecond laser with the original settings was ...

متن کامل

Opaque Bubble Layer Risk Factors in Femtosecond Laser-assisted LASIK.

PURPOSE To determine the characteristics and risk factors for occurrence of opaque bubble layer (OBL) during femtosecond laser-assisted flap creation for LASIK. METHODS One hundred ninety-eight eyes of 102 consecutive patients who underwent LASIK flap creation performed with the Alcon WaveLight FS200 laser (Alcon Laboratories, Inc., Fort Worth, TX) were retrospectively analyzed in a cohort st...

متن کامل

In vivo structural characteristics of the femtosecond LASIK-induced opaque bubble layers with ultrahigh-resolution SD-OCT.

The authors report in vivo morphology of opaque bubble layers with ultrahigh-resolution anterior-segment optical coherence tomography (UHR-OCT) in 3 patients. Two patients were operated on with a 30-kHz IntraLase femtosecond laser (Abbott Medical Optics, Abbott Park, IL) and one patient was operated on with a 500-kHz VisuMax femtosecond laser (Carl Zeiss Meditec, Jena, Germany). UHR-OCT images ...

متن کامل

Predictability of Corneal Flap Thickness in LASIK Using the WaveLight FS200 Femtosecond Laser

250 of 250 words) Purpose: To compare the intended versus resultant thickness of laser in situ keratomileusis (LASIK) flaps created with a new femtosecond laser (WaveLight FS200 Femtosecond Laser, WaveLight GmbH, Alcon Laboratories) and to report refractive outcomes at 3 months posttreatment. Setting: Wellington Eye Clinic, Dublin, Ireland Design: Retrospective consecutive case series Methods: ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2013