Final 5-year results of the TAXUS II trial: a randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for de novo coronary artery lesions.
نویسندگان
چکیده
BACKGROUND The TAXUS II trial was designed to evaluate the safety and efficacy of the commercialized slow-release (SR) and an investigation-only moderate-release (MR) polymer-based TAXUS paclitaxel-eluting stent compared with a bare-metal stent for the treatment of de novo coronary lesions. METHODS AND RESULTS This prospective, randomized, double-blind, controlled trial enrolled 536 patients in 2 consecutive cohorts to compare TAXUS SR (n=131) and TAXUS MR (n=135) with an identical but uncoated bare-metal stent control (n=270). The present analysis reports final 5-year clinical outcomes of TAXUS II. At 5 years, both TAXUS SR and MR showed superior outcomes compared with control. The 5-year rates of major adverse cardiac events were 27.6%, 20.4%, and 15.1% (P=0.01); rates of target-vessel revascularization were 22.5%, 16.6%, and 9.0% (P=0.004); and rates of target-lesion revascularization were 18.4%, 10.3%, and 4.5% (P<0.001) for the control, TAXUS SR, and TAXUS MR groups, respectively. The rates of all-cause death and myocardial infarction were low and similar between groups, with 2 stent thromboses with bare-metal stents compared with no event beyond 2 years with either of the TAXUS stents. CONCLUSIONS TAXUS II is the first large TAXUS trial to have reached 5-year follow-up. Both the SR and MR stents lowered the rates of target-vessel and target-lesion revascularization, which indicates their sustained efficacy. Furthermore, the low overall rates of all death, myocardial infarction, and stent thrombosis support the long-term safety of the TAXUS stent system.
منابع مشابه
Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions.
BACKGROUND Early clinical studies demonstrated the feasibility of local paclitaxel delivery in reducing restenosis after treatment of de novo coronary lesions in small patient populations. METHODS AND RESULTS We conducted a randomized, double-blind trial of 536 patients at 38 medical centers evaluating slow-release (SR) and moderate-release (MR) formulations of a polymer-based paclitaxel-elut...
متن کاملOne-Year Clinical Results With the Slow-Release, Polymer-Based, Paclitaxel-Eluting TAXUS Stent
Background—The safety and efficacy of the slow-release, polymer-based, paclitaxel-eluting stent after implantation in a broad cross section of de novo coronary lesions at 1 year are unknown. Methods and Results—In the TAXUS-IV trial, 1314 patients with single de novo coronary lesions 10 to 28 mm in length, with reference-vessel diameter 2.5 to 3.75 mm, coverable by a single study stent, were pr...
متن کاملOne-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent in patients with diabetes mellitus.
Background—The safety and efficacy of the slow-release, polymer-based, paclitaxel-eluting stent after implantation in a broad cross section of de novo coronary lesions at 1 year are unknown. Methods and Results—In the TAXUS-IV trial, 1314 patients with single de novo coronary lesions 10 to 28 mm in length, with reference-vessel diameter 2.5 to 3.75 mm, coverable by a single study stent, were pr...
متن کاملOutcomes of paclitaxel-eluting stent implantation in patients with stenosis of the left anterior descending coronary artery.
OBJECTIVES We sought to examine the efficacy of paclitaxel-eluting stent implantation in the left anterior descending coronary artery (LAD). BACKGROUND Restenosis and recurrent cardiac events after percutaneous intervention are more common for lesions in the LAD than other native coronary arteries, and often necessitate bypass surgery. Drug-eluting stents may improve the long-term prognosis o...
متن کاملOne-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trial.
BACKGROUND The safety and efficacy of the slow-release, polymer-based, paclitaxel-eluting stent after implantation in a broad cross section of de novo coronary lesions at 1 year are unknown. METHODS AND RESULTS In the TAXUS-IV trial, 1314 patients with single de novo coronary lesions 10 to 28 mm in length, with reference-vessel diameter 2.5 to 3.75 mm, coverable by a single study stent, were ...
متن کاملذخیره در منابع من