A translational couch technique for total body irradiation

نویسندگان

  • Mehrdad Sarfaraz
  • Cedric Yu
  • D. J. Chen
  • Leon Der
چکیده

We have constructed a computer controlled translational couch to administer total body irradiation reproducibly and safely. The system has replaced the previous stationary anterior-posterior technique in our institution and 30 plus patients have been treated with it so far. In this technique, patients comfortably lie on a couch in supine and prone positions and are transported slowly through a narrow beam with the gantry in an upright position. Dose to the patient is determined by the couch velocity that is calculated based on physical parameters such as patient's dimensions, beam geometry, and machine dose rate. In our design, the couch velocity is continuously updated to compensate for machine dose rate fluctuations. The translational couch technique provides better dose uniformity within the patient compared to fixed beam techniques, and allows a more precise shielding block placement for organs at risk. At the same time, it presents a special challenge for dosimetry calculations. A dosimetry parameter is introduced that converts the moving beam output to the fixed beam output factor. Based on this factor, a simple dosimetry calculation method has been developed that takes advantage of conventional dosimetry parameters, eliminating extensive dosimetry measurements. Multiple point dose measurements within a phantom confirmed the validity of the calculation method.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of different TBI techniques in terms of dose homogeneity

Introduction: Total body irradiation (TBI) is a form of external beam radiotherapy which is used in conjunction with chemotherapy with the purpose of immunosuppression. Since the target in TBI is the whole body so a uniform dose distribution throughout the entire body during TBI is necessary. As recommended by AAPM dose variation must be within ±10% of prescription dose. A revi...

متن کامل

Treatment couch positioning uncertainties using an EPID-based method

Introduction: The accuracy of patient positioning plays an important role for radiotherapy tasks. Short and automated Quality Assurance (QA) programs are becoming one of the challenging tasks in Radiotherapy. The current study, investigates the accuracy of treatment couch positioning with four degrees of freedom (4DoF) using a fast and accurate method based on an image acquired...

متن کامل

Dose Evaluation in the Movement Couch of the Total Body Irradiation Technique Using Semiconducting Diodes and Thermoluminescence Detectors

Aim: The present study is devoted to evaluate the dose at different points defined on the patient using semiconducting diodes and thermoluminescence detectors to excute verification of the dose at these relevant points calculated by means of movement couch and Beam Zones method of the Total Body Irradiation Technique. Material and Method: In this technique; 11 patients using 6 MV X-ray in the (...

متن کامل

AbstractID: 13162 Title: 3D heterogeneous dose distributions for Total Body Irradiation Patients

ID: 13162 Title: 3D heterogeneous dose distributions for Total Body Irradiation Patients Purpose: One major objective of Total Body Irradiation (TBI) treatments is to deliver a uniform dose in the entire body of the patient. Looking at 3D dose distributions for constant speed (CteSpeed) and variable speed (VarSpeed) translating couch TBI treatments, dose uniformity and the effect of body hetero...

متن کامل

Evaluation of dose uniformity in total body irradiation (TBI) technique in parallel-opposed anterior/posterior geometry using Varian Clinac 2100C/D linear accelerator

Background: Total body irradiation (TBI) is a technique that is commonly used as a part of the patient conditioning regimen before the bone marrow transplant (BMT). The purpose of this study is to introduce and implement a reasonable TBI technique on the human-like phantom in Imam Khomeini Hospital in Tehran. Methods: The present experimental study was conducted from October 2016 to November 2...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2  شماره 

صفحات  -

تاریخ انتشار 2001