Medium-term outcome of titanium nitride, mobile bearing total knee replacement.
نویسندگان
چکیده
This retrospective study examined the medium term outcome of 305 Titanium Nitride (TiN) Mobile bearing total knee. At ten years after the start of the study, there was a significant improvement (p < 0.0001) in the post-operative knee scores. The ten year survival with revision for any reason as the end point was 95.1% (95% CI 92.4 to 97.8). The ten year survival rate with revision for aseptic loosening as the end point was 99.1% (95% CI 97.8 to 100%). A total of 15 knees (4.9%, overall) required revision. No cases were revised due to sepsis. Based on these results, the surface Titanium Nitride ceramic, mobile-bearing total knee replacement has proved to be a reliable implant at 10 years when used in primary knee replacement.
منابع مشابه
Evaluation of 1031 primary titanium nitride coated mobile bearing total knee arthroplasties in an orthopedic clinic
AIM To evaluate the influence of the titanium nitride (TiN) coating on the results of a total knee arthroplasty (TKA). METHODS A total of 910 patients (338 men; 572 woman), with a mean age of 65 (range 36-94) undergoing 1031 primary TKAs were assessed. Clinical evaluation and patient-reported outcomes were gathered one year after surgery. The questionnaires included the Knee injury and Osteoa...
متن کاملOne Year After Navigated Total Knee Replacement, No Clinically Relevant Difference Found Between Fixed Bearing and Mobile Bearing Knee Replacement in a Double-Blind Randomized Controlled Trial
BACKGROUND The mobile bearing designs have not yet been shown to improve clinical outcome of total knee arthroplasty (TKA). In this prospective randomized study, we compared the short-term clinical results of a mobile bearing implant with those of the fixed bearing version of the same implant. METHODS We randomized 100 knees into two double-blind groups who received either the fixed (FB, 52 k...
متن کاملFixed or mobile-bearing total knee arthroplasty
Fixed and mobile-bearing in total knee arthroplasty are still discussed controversially. In this article, biomechanical and clinical aspects in both fixed and mobile-bearing designs were reviewed. In biomechanical aspect, the mobile-bearing design has proved to provide less tibiofemoral contact stresses under tibiofemoral malalignment conditions. It also provides less wear rate in in-vitro simu...
متن کاملAbrasion resistance of oxidized zirconium in comparison with CoCrMo and titanium nitride coatings for artificial knee joints.
Most total knee replacement joints consist of a metal femoral component made from a cobalt-chromium- molybdenum (CoCrMo)-alloy and a tibial component with an ultrahigh molecular weight polyethylene (UHMWPE) bearing surface. Wear of the UHMWPE remains the primary disadvantage of these implants. The allergic potential ascribed to CoCrMo-alloys is a further concern. Other metallic alloys with and ...
متن کاملTotal knee arthroplasty with mobile tibial weight-bearing: clinical evaluation after a minimum of five years of postoperative follow-up☆
OBJECTIVE To evaluate the medium and long term results from total knee arthroplasty with rotating tibial weight-bearing. METHODS Between January 2000 and July 2007, 162 patients underwent total knee arthroplasty with mobile tibial weight-bearing. Among these, 96 were evaluated in a previous study with a mean follow-up of 4 years. In the present study, the same group was invited back for reass...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Acta orthopaedica Belgica
دوره 80 2 شماره
صفحات -
تاریخ انتشار 2014