Comparison of Perkins, Tono-Pen and Schiøtz tonometers in paediatric patients under general anaesthesia.
نویسندگان
چکیده
This cross-sectional study aimed to investigate if the Tono-Pen, Schiøtz and Perkins tonometers could be used interchangeably in general practice for measuring elevated intraocular pressure. A total of 74 eyes of 37 paediatric patients under general anaesthesia were checked with all 3 tonometers. All of the tonometers gave significantly different measurements from each other. However, with a mean difference of 1.4 mmHg and 95% limits of agreement of -5.7 to +8.6, the greatest agreement was between the Perkins and Tono-Pen tonometers. The Perkins tonometer is a hand-held variant of the Goldmann tonometer (the gold standard for intraocular pressure measures). Therefore the Tono-Pen with its ease of use and safety could be a reliable device for use in general practice.
منابع مشابه
Clinical comparison of the ProTon and Tono-Pen tonometers with the Goldmann applanation tonometer.
A clinical evaluation of a new electron ProTon tonometer was performed comparing the values of intraocular pressure (IOP) measured using this instrument with those determined by a similar instrument, Tono-Pen XL, and by Goldmann applanation tonometry. The mean IOP measured in 106 eyes with the ProTon tonometer was not significantly different from that determined with Goldmann applanation, while...
متن کاملClinical classification and medical treatment options in childhood glaucoma.
IOP is the most variable diagnostic as it could be due to multiple factors (tonometer type, ocular movements, patient cooperation, anesthesia or corneal condition). Reference tonometry is Goldmann’s tonometer, although the Perkins tonometer (Haag-Streit AG, Bern, Switzerland) was utilized on multiple occasions as it enables IOP readings in supine position and with uncooperative patients. Other ...
متن کاملComparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL
OBJECTIVE WE AIMED TO COMPARE INTRAOCULAR PRESSURE (IOP) MEASUREMENTS VIA THREE DIFFERENT TONOMETERS: the Goldmann applanation tonometer (GAT), the Tono-Pen(®) XL (TPXL), and a non-contact airpuff tonometer (NCT). METHODS This was a cross-sectional study of 200 eyes from 200 patients. Right eyes of all patients were included in this study. IOP was measured via GAT, NCT, and TPXL by three phys...
متن کاملIntraocular Pressure Measurements by Three Different Tonometers in Children with Aphakic Glaucoma and a Thick Cornea
Background: To evaluate the agreement in intraocular pressure (IOP) measurements by Ocular Response Analyzer (ORA) and Tono-Pen XL (TXL) with the Goldmann Applanation Tonometer (GAT) and to examine corneal biomechanical properties in aphakic glaucoma patients with a central corneal thickness (CCT) >600 µ. Methods: Thirty-six eyes of aphakic glaucoma patients (group 1) and 40 eyes of normal chil...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Eastern Mediterranean health journal = La revue de sante de la Mediterranee orientale = al-Majallah al-sihhiyah li-sharq al-mutawassit
دوره 14 6 شماره
صفحات -
تاریخ انتشار 2008