Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197.
نویسندگان
چکیده
BACKGROUND Accurate and reliable blood glucose (BG) measurements require that different test strip lots of the same BG monitoring system provide comparable measurement results. Only a small number of studies addressing this question have been published. METHODS In this study, four test strip lots for each of five different BG systems [Accu-Chek® Aviva (system A), FreeStyle Lite® (system B), GlucoCheck XL (system C), Pura™/mylife™ Pura (system D), and OneTouch® Verio™ Pro (system E)] were evaluated with procedures according to DIN EN ISO 15197:2003. The BG system measurement results were compared with the manufacturer's measurement procedure (glucose oxidase or hexokinase method). Relative bias according to Bland and Altman and system accuracy according to ISO 15197 were analyzed. A BG system consists of the BG meter itself and the test strips. RESULTS The maximum lot-to-lot difference between any two of the four evaluated test strip lots per BG system was 1.0% for system E, 2.1% for system A, 3.1% for system C, 6.9% for system B, and 13.0% for system D. Only two systems (systems A and B) fulfill the criteria of DIN EN ISO 15197:2003 with each test strip lot. CONCLUSIONS Considerable lot-to-lot variability between test strip lots of the same BG system was found. These variations add to other sources of inaccuracy with the specific BG system. Manufacturers should regularly and effectively check the accuracy of their BG meters and test strips even between different test strip lots to minimize risk of false treatment decisions.
منابع مشابه
System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197.
BACKGROUND The accuracy of systems for self-monitoring of blood glucose is important, as reliable measurement results are a prerequisite for therapeutic decisions. METHODS This system accuracy evaluation study was performed according to DIN EN ISO 15197:2003 for 43 Conformité Européenne (CE)-labeled blood glucose (BG) monitoring systems. Measurement results of each system were compared with r...
متن کاملSystem accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197.
BACKGROUND Blood glucose (BG) monitoring systems enable diabetes patients to effectively control and adjust their therapy. BG monitoring systems with a Conformité Européenne (CE) label should meet the standard DIN EN ISO 15197:2003: > or =95% of the BG results shall fall within +/-15 mg/dL of the reference method at BG concentrations <75 mg/dL and within +/-20% at BG concentrations > or =75 mg/...
متن کاملA new test strip technology platform for self-monitoring of blood glucose.
In the management of diabetes, accuracy of devices used for self-monitoring of blood glucose (SMBG) is critical because SMBG results can affect patient diabetes-related health outcomes. A new blood glucose monitoring system (BGMS) platform has been developed that is based on the new CONTOUR® NEXT (CN) test strip. This BGMS platform uses a proprietary electron mediator and algorithm to minimize ...
متن کاملAccuracy of Four Blood Glucose Monitoring Systems
Background: The minimum acceptable accuracy for results produced by a blood glucose monitoring system according to the current version of ISO 15197, created in 2003, is: 95% of the individual glucose results shall fall within ±15 mg/dL (0.83 mmol/L) of the results of the manufacturer’s measurement procedure at glucose concentrations < 75 mg/dL (< 4.2 mmol/L) and within ±20% at glucose concentra...
متن کاملPerformance variability of seven commonly used self-monitoring of blood glucose systems: clinical considerations for patients and providers.
BACKGROUND Blood glucose data are frequently used in clinical decision making, thus it is critical that self-monitoring of blood glucose (SMBG) systems consistently provide accurate results. Concerns about SMBG accuracy have prompted the development of newly proposed International Organization for Standardization (ISO) standards: ≥ 95% of individual glucose results shall fall within ± 15 mg/dl ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of diabetes science and technology
دوره 6 5 شماره
صفحات -
تاریخ انتشار 2012