Non-invasive continuous core temperature measurement by zero heat flux.

نویسندگان

  • L P J Teunissen
  • J Klewer
  • A de Haan
  • J J de Koning
  • H A M Daanen
چکیده

Reliable continuous core temperature measurement is of major importance for monitoring patients. The zero heat flux method (ZHF) can potentially fulfil the requirements of non-invasiveness, reliability and short delay time that current measurement methods lack. The purpose of this study was to determine the performance of a new ZHF device on the forehead regarding these issues. Seven healthy subjects performed a protocol of 10 min rest, 30 min submaximal exercise (average temperature increase about 1.5 °C) and 10 min passive recovery in ambient conditions of 35 °C and 50% relative humidity. ZHF temperature (T(zhf)) was compared to oesophageal (T(es)) and rectal (T(re)) temperature. ΔT(zhf)-T(es) had an average bias ± standard deviation of 0.17 ± 0.19 °C in rest, -0.05 ± 0.18 °C during exercise and -0.01 ± 0.20 °C during recovery, the latter two being not significant. The 95% limits of agreement ranged from -0.40 to 0.40 °C and T(zhf) had hardly any delay compared to T(es). T(re) showed a substantial delay and deviation from T(es) when core temperature changed rapidly. Results indicate that the studied ZHF sensor tracks T(es) very well in hot and stable ambient conditions and may be a promising alternative for reliable non-invasive continuous core temperature measurement in hospital.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Continuous Core Body Temperature Estimation via SURFACE Temperature Measurements using Wearable Sensors - Is it Feasible?

Core body temperature is an important indicator of well being of humans. The commonly used methods and sites of measurement do not lend well for continuous measurement at home. However, continuous monitoring using non-invasive, small, low cost sensors could have many applications like detection of hypothermia and fever in low birth weight neonates in rural settings. We investigate the feasibili...

متن کامل

Novel Zero-Heat-Flux Deep Body Temperature Measurement in Lower Extremity Vascular and Cardiac Surgery.

OBJECTIVE The aim of this study was to compare deep body temperature obtained using a novel noninvasive continuous zero-heat-flux temperature measurement system with core temperatures obtained using conventional methods. DESIGN A prospective, observational study. SETTING Operating room of a university hospital. PARTICIPANTS The study comprised 15 patients undergoing vascular surgery of th...

متن کامل

An evaluation of a zero-heat-flux cutaneous thermometer in cardiac surgical patients.

BACKGROUND Although core temperature can be measured invasively, there are currently no widely available, reliable, noninvasive thermometers for its measurement. We thus compared a prototype zero-heat-flux thermometer with simultaneous measurements from a pulmonary artery catheter. Specifically, we tested the hypothesis that zero-heat-flux temperatures are sufficiently accurate for routine clin...

متن کامل

Heat Transfer in Semitransparent Medium Caused by Laser Pulse

In this paper, the combination of conduction with radiation into a semitransparent medium which includes absorption, emission and scattering has been investigated. In order to Study the conduction in medium, the Non-Fourier heat conduction has been applied. In this model there is a time delay between heat flux and temperature gradient. Also, in contrast with Fourier heat conduction, the speed o...

متن کامل

Non-invasive, transient determination of the core temperature of a heat-generating solid body

While temperature on the surface of a heat-generating solid body can be easily measured using a variety of methods, very few techniques exist for non-invasively measuring the temperature inside the solid body as a function of time. Measurement of internal temperature is very desirable since measurement of just the surface temperature gives no indication of temperature inside the body, and syste...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Physiological measurement

دوره 32 5  شماره 

صفحات  -

تاریخ انتشار 2011