Comparative study between I-gel, a new supraglottic airway device, and classical laryngeal mask airway in anesthetized spontaneously ventilated patients
نویسندگان
چکیده
OBJECTIVE To compare two different supraglottic airway devices, the laryngeal mask airway (LMA) and the I-gel, regarding easiness of insertion of the device, leak pressure, gastric insufflation, end tidal CO2, oxygen saturation, hemodynamic and postoperative complications in anesthetized, spontaneously ventilated adult patients performing different non-emergency surgical procedures. MATERIALS AND METHODS The study was carried out as a prospective, randomized, clinical trial among 80 patients who underwent different surgical procedures under general anesthesia with spontaneous ventilation in supine position. They were equally randomized into two groups: I-gel and LMA groups. Both the devices were compared with regard to heart rate, arterial BP, SPO2, end-tidal CO2, number and duration of insertion attempts, incidence of gastric insufflation, leak pressure and airway assessment after removal of the device. RESULTS No statistically significant difference was reported between both the groups, regarding heart rate, arterial BP, SPO2 and end-tidal CO2. The mean duration of insertion attempts was 15.6 +/- 4.9 seconds in the I-gel group, while it was 26.2 +/- 17.7 seconds in the LMA group. The difference between both the groups regarding duration of insertion attempts was statistically significant (P 0.0023*), while the number of insertion attempts was statistically insignificant between both the study groups (P > 0.05). Leak pressure was(25.6 +/- 4.9 versus 21.2 +/- 7.7 0.016* cmH2O) significantly higher among studied patients of the I-gel group and incidence of gastric insufflation was significantly more with LMA 9 (22.5%) versus 2 (5%) 0.016* in I-gel group. CONCLUSION Both LMA and I-gel do not cause any significant alteration in the hemodynamic status of the patients, end tidal CO2, and SPO2. The postoperative complications were not significantly different except nusea and vomiting was statistically significant higher in LMAgroup(P 0.032). among both LMA and I-gel patients. Insertion of I-gel was significantly easier and more rapid than insertion of LMA. Leak pressure was significantly higher with I-gel than LMA and thus incidence of gastric insufflation was significantly lower with I-gel.
منابع مشابه
Comparative evaluation of LMA-Supreme and i-gel supraglottic airway devices with endotracheal intubation during surgical correction of traumatic orbital injuries
Aims. The aim of the study was to compare endotracheal intubation with i-gel and LMA (laryngeal mask airway) supreme supraglottic airway devices in orbital osteosynthesis surgery. Methods. 134 patients undergoing orbital plasty after fracture were enrolled in our study and were randomized into three groups: 43 in the endotracheal intubation group, 42 in the LMA-Supreme group, and 49 patients w...
متن کاملA Comparison of Three Different Supraglottic Airway Devices in Neonatal Airway Training during Resuscitation Simulation
Objective: The aim of this study was to compare the success rates for insertion using the laryngeal mask, laryngeal tube, and I-gel supraglottic airway devices in a neonatal airway manikin during a resuscitation simulation course. Methods: Three groups of health care professionals were given a brief supervised training in using the devices. For each participant the positioning of each device wa...
متن کاملA comparative Study of Tracheal Intubation through i-gel® and Intubating
In this prospective randomized study, success rate of blind tracheal intubation through two different supraglottic airway devices viz. i-gel ® and the intubating laryngeal mask airway (ILMA)® was compared a using conventional polyvinylchloride endotracheal tubes. Eighty patients undergoing elective surgery under general anaesthesia were randomized in two groups comprising of 40 patients each to...
متن کاملComparison of the i-gel™ and the Laryngeal Mask Airway Classic™ in terms of clinical performance.
PURPOSE The i-gel™ is one of the second generation supraglottic airway devices. Our study was designed to compare the i-gel and the Laryngeal Mask Airway Classic™ with respect to the clinical performance. METHODS We compared the performance of the i-gel with that of the Laryngeal Mask Airway Classic in 120 patients undergoing urologic surgery during general anesthesia without muscle relaxant ...
متن کاملThe i-gel Supraglottic Airway as a Conduit for Fibreoptic Tracheal Intubation - A Randomized Comparison with the Single-use Intubating Laryngeal Mask Airway and CTrach Laryngeal Mask in Patients with Predicted Difficult Laryngoscopy.
Fibreoptic intubation through a supraglottic airway is an alternative plan for airway management in difficult or failed laryngoscopy. The aim of this study was to compare three supraglottic airways as conduits in patients with at least one predictor for difficult laryngoscopy. The i-gel was compared with the single-use intubating laryngeal mask airway (sILMA) and CTrach laryngeal mask in 120 ad...