OBJECTIVES This study sought to examine the safety and efficacy of laser-assisted lead extraction and the indications, outcomes, and risk factors in a large series of consecutive patients. BACKGROUND The need for lead extraction has been increasing in direct relationship to the increased numbers of cardiovascular implantable electronic devices. METHODS Consecutive patients undergoing transvenous laser-assisted lead extraction at 13 centers were included. RESULTS Between January 2004 and December 2007, 1,449 consecutive patients underwent laser-assisted lead extraction of 2,405 leads (20 to 270 procedures/site). Median implantation duration was 82.1 months (0.4 to 356.8 months). Leads were completely removed 96.5% of the time, with a 97.7% clinical success rate whereby clinical goals associated with the indication for lead removal were achieved. Failure to achieve clinical success was associated with body mass index <25 kg/m(2) and low extraction volume centers. Procedural failure was higher in leads implanted for >10 years and when performed in low volume centers. Major adverse events in 20 patients were directly related to the procedure (1.4%) including 4 deaths (0.28%). Major adverse effects were associated with patients with a body mass index <25 kg/m(2). Overall all-cause in-hospital mortality was 1.86%; 4.3% when associated with endocarditis, 7.9% when associated with endocarditis and diabetes, and 12.4% when associated with endocarditis and creatinine > or =2.0. Indicators of all-cause in-hospital mortality were pocket infections, device-related endocarditis, diabetes, and creatinine > or =2.0. CONCLUSIONS Lead extraction employing laser sheaths is highly successful with a low procedural complication rate. Total mortality is substantially increased with pocket infections or device-related endocarditis, particularly in the setting of diabetes, renal insufficiency, or body mass index <25 kg/m(2). Centers with smaller case volumes tended to have a lower rate of successful extraction.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Leaching of heavy metal from native plants by chemical leachants

Plant biomass harvested after heavy-metal phytoremediation must be considered as a hazardous waste that should be contained or treated appropriately before disposal or reuse. This study provides an evaluation of extractants for the removal of lead from Abutilon Theophrastus biomass. The research was carried out a leaching study to determine the lead-extraction efficiency of the different leacha...

متن کامل

Lead extraction for device related infections: a single-centre experience.

AIMS We report a single-centre experience of extraction of infected pacing and ICD leads. METHODS Retrospective study of consecutive lead extractions for infection. Lead extraction was by traction, or, if unsuccessful, a laser sheath was used. All procedures were performed in the operating room. RESULTS Leads were extracted in 82 patients including 66 patients (80%) with infection occurring...

متن کامل

A single-centre experience of over one thousand lead extractions

AIMS The aim of the study was to present a single-centre experience of pacemaker and implantable cardioverter defibrillator (ICD) lead extraction using different methods, mainly laser-assisted extraction. METHODS AND RESULTS Data from 1032 leads and 647 procedures were gathered. A step-by-step approach using different techniques while performing an ongoing risk-benefit analysis was used. The ...

متن کامل

Evaluation of two different extraction techniques on determination of 〖Pb〗^(+2) contents of wastes of dye manufacturing industries

Extraction of trace amounts of toxic heavy metals from complicated environmental matrixes is a  challenging task. The aim of this investigation is the comparison of the extraction efficiencies and other related experimental variables ( chemicals, effect of pH, potential interfering ions) on the determination of inorganic lead ( ) which is present in waste water of dye industries. Compared techn...

متن کامل

Medication Errors in Administration of Chemotherapeutic Agents: an Observational Study

Chemotherapy medication errors may lead to potentially harmful consequences while most of them could be preventable. This study aims to determine the incidence and type of drug handling and administration errors among the nurses and to identify possible contributing factors. Setting of the study was a teaching hospital affiliated to Shahid Beheshti University of Medic...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of the American College of Cardiology

دوره 55 6  شماره 

صفحات  -

تاریخ انتشار 2010