Evaluation of Wall Correction Factor of INER's Air-Kerma Primary Standard Chamber and Dose Variation by Source Displacement for HDR 192Ir Brachytherapy

نویسندگان

  • J. H. Lee
  • J. N. Wang
  • T. T. Huang
  • S. H. Su
  • B. J. Chang
  • C. H. Su
  • S. M. Hsu
چکیده

The aim of the present study was to estimate the wall effect of the self-made spherical graphite-walled cavity chamber with the Monte Carlo method for establishing the air-kerma primary standard of high-dose-rate (HDR) ¹⁹²Ir brachytherapy sources at the Institute of Nuclear Energy Research (INER, Taiwan). The Monte Carlo method established in this paper was also employed to respectively simulate wall correction factors of the ¹⁹²Ir air-kerma standard chambers used at the National Institute of Standards and Technology (NIST, USA) and the National Physical Laboratory (NPL, UK) for comparisons and verification. The chamber wall correction calculation results will be incorporated into INER's HDR ¹⁹²Ir primary standard in the future. For the brachytherapy treatment in the esophagus or in the bronchi, the position of the isotope may have displacement in the cavity. Thus the delivered dose would differ from the prescribed dose in the treatment plan. We also tried assessing dose distribution due to the position displacement of HDR ¹⁹²Ir brachytherapy source in a phantom with a central cavity by the Monte Carlo method. The calculated results could offer a clinical reference for the brachytherapy within the human organs with cavity.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy

Background: The effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted Standards. Secondary Standard Dosimetry Laboratory (SSDL) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192Ir . This work has been carried out to calibrate the high dose rat...

متن کامل

A practical implementation of the 2010 IPEM high dose rate brachytherapy code of practice for the calibration of 192Ir sources.

This paper details a practical method for deriving the reference air kerma rate calibration coefficient for Farmer NE2571 chambers using the U.K. Institute of Physics and Engineering in Medicine (IPEM) code of practice for the determination of the reference air kerma rate for HDR (192)Ir brachytherapy sources based on the National Physical Laboratory (NPL) air kerma standard. The reference air ...

متن کامل

In-air calibration of an HDR 192Ir brachytherapy source using therapy ion chambers.

The Gammamed Plus 192Ir high dose rate brachytherapy sources were calibrated using the therapy level ionization chambers (0.1 and 0.6 cc) and the well-type chamber. The aim of the present study was to assess the accuracy and suitability of use of the therapy level chambers for in-air calibration of brachytherapy sources in routine clinical practice. In a calibration procedure using therapy ion ...

متن کامل

Independent assessment of source transit time for the BEBIG SagiNova® high dose rate brachytherapy afterloader

Introduction: The dwell time and transit time components contribute to the overall delivered dose to patients in high dose rate (HDR) brachytherapy treatments. The transit time results from source entry and exit as well as source movements between dwell positions. It depends on various parameters such as the source speed profile, source indexer distance to dwell position, and ...

متن کامل

Comparison of Calibration Methods for an Iridium 192 Source (hdr) and Application to the Absorbed Dose to Water Determination

The aim of this study was to compare the high dose rate (HDR) iridium 192 source calibration in terms of reference air kerma rate with two types of ionisation chambers (a thimble chamber and a well-type chamber). Dose measurements in water have been also performed with thimble chamber and thermoluminescent dosimeters (TLD) in order to validate the absorbed dose to water calculation provided by ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2013  شماره 

صفحات  -

تاریخ انتشار 2013