Noncontact diffuse correlation spectroscopy for noninvasive deep tissue blood flow measurement.
نویسندگان
چکیده
A noncontact diffuse correlation spectroscopy (DCS) probe has been developed using two separated optical paths for the source and detector. This unique design avoids the interference between the source and detector and allows large source-detector separations for deep tissue blood flow measurements. The noncontact probe has been calibrated against a contact probe in a tissue-like phantom solution and human muscle tissues; flow changes concurrently measured by the two probes are highly correlated in both phantom (R(2)=0.89, p<10(-5)) and real-tissue (R(2)=0.77, p<10(-5), n=9) tests. The noncontact DCS holds promise for measuring blood flow in vulnerable (e.g., pressure ulcer) and soft (e.g., breast) tissues without distorting tissue hemodynamic properties.
منابع مشابه
Simultaneous measurement of deep tissue blood flow and oxygenation using noncontact diffuse correlation spectroscopy flow-oximeter
We report a novel noncontact diffuse correlation spectroscopy flow-oximeter for simultaneous quantification of relative changes in tissue blood flow (rBF) and oxygenation (Δ[oxygenation]). The noncontact probe was compared against a contact probe in tissue-like phantoms and forearm muscles (n = 10), and the dynamic trends in both rBF and Δ[oxygenation] were found to be highly correlated. Howeve...
متن کاملA Novel Noncontact Diffuse Correlation Spectroscopy Device for Assessing Blood Flow in Mastectomy Skin Flaps: A Prospective Study in Patients Undergoing Prosthesis-Based Reconstruction.
A new advanced technology, noncontact diffuse correlation spectroscopy, has been recently developed for the measurement of tissue blood flow through analyzing the motions of red blood cells in deep tissues. This technology is portable, inexpensive, and noninvasive, and can measure up to 1.5-cm tissue depth. In this prospective study, the authors aimed to explore the use of this novel device in ...
متن کاملThree-dimensional flow contrast imaging of deep tissue using noncontact diffuse correlation tomography.
This study extended our recently developed noncontact diffuse correlation spectroscopy flowmetry system into noncontact diffuse correlation tomography (ncDCT) for three-dimensional (3-D) flow imaging of deep tissue. A linear array of 15 photodetectors and two laser sources connected to a mobile lens-focusing system enabled automatic and noncontact scanning of flow in a region of interest. These...
متن کاملMultidistance diffuse correlation spectroscopy for simultaneous estimation of blood flow index and optical properties.
Traditionally, diffuse correlation spectroscopy (DCS) measures microvascular blood flow by fitting a physical model to the measurement of the intensity autocorrelation function from a single source-detector pair. This analysis relies on the accurate knowledge of the optical properties, absorption, and reduced scattering coefficients of the medium. Therefore, DCS is often deployed together with ...
متن کاملNoninvasively measuring the hemodynamic effects of massage on skeletal muscle: a novel hybrid near-infrared diffuse optical instrument.
Increase in tissue blood flow is one of the most acknowledged potential effects of massage; however, actual research studies examining this phenomenon are inconsistent and inconclusive. One possible reason for continued uncertainty regarding this topic is methodology, specifically how tissue blood flow is measured because limitations exist in previously utilized technologies. Near-infrared spec...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of biomedical optics
دوره 17 1 شماره
صفحات -
تاریخ انتشار 2012