Quantifying cross-scatter contamination in biplane fluoroscopy motion analysis systems.

نویسندگان

  • Janelle A Cross
  • Ben McHenry
  • Taly Gilat Schmidt
چکیده

Biplane fluoroscopy is used for dynamic in vivo three-dimensional motion analysis of various joints of the body. Cross-scatter between the two fluoroscopy systems may limit tracking accuracy. This study measured the magnitude and effects of cross-scatter in biplane fluoroscopic images. Four cylindrical phantoms of 4-, 6-, 8-, and 10-in. diameter were imaged at varying kVp levels to determine the cross-scatter fraction and contrast-to-noise ratio (CNR). Monte Carlo simulations quantified the effect of the gantry angle on the cross-scatter fraction. A cadaver foot with implanted beads was also imaged. The effect of cross-scatter on marker-based tracking accuracy was investigated. Results demonstrated that the cross-scatter fraction varied from 0.15 for the 4-in. cylinder to 0.89 for the 10-in. cylinder when averaged across kVp. The average change in CNR due to cross-scatter ranged from 5% to 36% CNR decreases for the 4- and 10-in. cylinders, respectively. In simulations, the cross-scatter fraction increased with the gantry angle for the 8- and 10-in. cylinders. Cross-scatter significantly increased static-tracking error by 15%, 25%, and 38% for the 6-, 8-, and 10-in. phantoms, respectively, with no significant effect for the foot specimen. The results demonstrated submillimeter marker-based tracking for a range of phantom sizes, despite cross-scatter degradation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Marker-based validation of a biplane fluoroscopy system for quantifying foot kinematics.

INTRODUCTION Radiostereometric analysis has demonstrated its capacity to track precise motion of the bones within a subject during motion. Existing devices for imaging the body in two planes are often custom built systems; we present here the design and marker-based validation of a system that has been optimized to image the foot during gait. METHODS Mechanical modifications were made to pair...

متن کامل

Accuracy of a contour-based biplane fluoroscopy technique for tracking knee joint kinematics of different speeds.

While measuring knee motion in all six degrees of freedom is important for understanding and treating orthopaedic knee pathologies, traditional motion capture techniques lack the required accuracy. A variety of model-based biplane fluoroscopy techniques have been developed with sub-millimeter accuracy. However, no studies have statistically evaluated the consistency of the accuracy across motio...

متن کامل

Preliminary model-based validation of a biplane fluoroscopy system

Background Biplane fluoroscopy can directly track the motion of bones and therefore measure joint kinematics. Our prior markerbased work has demonstrated the ability of our system to accurately and precisely track the motion of known objects (i.e., tantalum beads) [1]. In this study, we present the preliminary bone-based validation of our system by tracking the bones of the foot from cadaveric ...

متن کامل

Dynamic Three-Dimensional Analysis of Lachman Test for Anterior Cruciate Ligament Insufficiency: Analysis of Anteroposterior Motion of the Medial and Lateral Femoral Epicondyles

PURPOSE To evaluate dynamic three-dimensional (3D) kinematic properties of the anterior cruciate ligament (ACL)-insufficient knees and healthy contralateral knees in awake patients during the Lachman test using biplane fluoroscopy. MATERIALS AND METHODS Ten patients with unilateral ACL-insufficient knees and healthy contralateral knees were enrolled in this study. Each patient underwent the L...

متن کامل

3-D Respiratory Motion Compensation during EP Procedures by Image-Based 3-D Lasso Catheter Model Generation and Tracking

Radio-frequency catheter ablation of the pulmonary veins attached to the left atrium is usually carried out under fluoroscopy guidance. Two-dimensional X-ray navigation may involve overlay images derived from a static pre-operative 3-D volumetric data set to add anatomical details. However, respiratory motion may impair the utility of static overlay images for catheter navigation. We developed ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of medical imaging

دوره 2 4  شماره 

صفحات  -

تاریخ انتشار 2015