SPECT using a specially designed cone beam collimator.
نویسندگان
چکیده
A specially designed high resolution converging collimator having a focal length of 50 cm has been evaluated for cone beam single photon emission computed tomography (SPECT). The focal region was investigated by imaging a point source placed at the expected focal point and along the central ray of the collimator in front of and behind the focal point. Technetium-99m point source sensitivities measured in air at 5, 10, 15, and 20 cm from the collimator surface are 4.2, 5.5, 7.3, and 10.5 cts.sec-1.microCi-1 when used with a single camera SPECT system. A commercially available parallel hole collimator, with similar resolution characteristics has a measured sensitivity of 3.3 cts.sec-1.microCi-1. Volume sensitivities of 9,780 and 4,945 (cts.sec-1)/(microCi.ml-1) were measured for the cone beam and parallel hole collimators, respectively, using a 17-cm-diameter spherical source. Reconstructed spatial resolution (FWHM) on the axis-of-rotation ranged between 10 and 11 mm for both collimators when the radius of rotation was equal to 15 cm. Using equal acquisition times SPECT images of phantoms scanned with the cone beam collimator were visually improved compared with images acquired using the parallel hole collimator. These results demonstrate that a factor of 2 improvement in volume sensitivity can be demonstrated with a cone beam collimator compared with a commercially available parallel hole collimator. Further improvements are possible using shorter focal lengths, astigmatic focusing, and larger field of view cameras.
منابع مشابه
Pinhole SPECT vs. Cone-Beam SPECT
In single photon emission computed tomography (SPECT), pinhole and cone-beam collimators are used for small object imaging. It is accepted that a pinhole collimator should be used if the object is very small, and the cone-beam collimator should be used if the object is larger. This paper presents criteria to determine which collimation geometry is more advantageous in terms of spatial resolutio...
متن کاملA Generalized Model for 3D Compton Scatter in Single Photon Emission Computed Tomography Using Slice-by-Slice Blurring and Scatter Projection Rebinning
The modeling of 3D Compton scatter in SPECT is complicated by the use of a variety of collimator geometry. To date, effective and efficient modeling of 3D Compton scatter in parallel-beam SPECT has been made possible by Frey and Tsui [1], Zeng et al [2], and Bai et al [3]. For fan-beam and cone-beam SPECT, the model in [3] can effectively model first order 3D Compton scatter. But for varying-fo...
متن کاملPerformance of myocardial perfusion imaging using multi-focus fan beam collimator with resolution recovery reconstruction in a comparison with conventional SPECT
Objective: IQSPECT is an advanced high-speed SPECT modality for performing myocardial perfusion imaging (MPI), which uses a multi-focus fan beam collimator with resolution recovery reconstruction. The aim of this study was to compare IQSPECT compared with conventional SPECT interms of performance based on standard clinical protocols. In addition, we examined the concordance between convention...
متن کاملA skew-slit collimator for small-animal SPECT.
UNLABELLED The main objective of this work was the development of a skew-slit collimator for small-animal SPECT to replace the state-of-the-art multipinhole collimator. METHODS A pinhole forms a cone-beam imaging geometry. If the collimator rotates around the object in a circular orbit, the projection measurements acquired by the cone-beam imaging geometry are incomplete and not enough for ar...
متن کاملEvaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT Images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
Objective(s): The purpose of this study was to examine the optimal reconstruction parameters for brain dopamine transporter SPECT images obtained with a fan beam collimator and compare the results with those obtained by using parallel-hole collimators.Methods: Data acquisition was performed using two SPECT/CT devices, namely a Symbia T6 and an Infinia Hawkeye 4 (device A and B) equipped with fa...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of nuclear medicine : official publication, Society of Nuclear Medicine
دوره 29 8 شماره
صفحات -
تاریخ انتشار 1988