Multi-dimensional dosimetric verification of stereotactic radiotherapy for uveal melanoma using radiochromic EBT film.
نویسندگان
چکیده
Since 1997, linac based stereotactic radiotherapy (SRT) of uveal melanoma has been continuously developed at the Department of Radiotherapy, Medical University Vienna. The aim of the present study was (i) to test a new type of radiochromic film (Gafchromic EBT) for dosimetric verification of class solutions for these treatments and (ii) to verify treatment plan acceptance criteria, which are based on gamma values statisitcs. An EPSON Expression 1680 Pro flat bed scanner was utilized for film reading. To establish a calibration curve, films were cut in squares of 2 x 2 cm2, positioned at 5 cm depth in a solid water phantom and were irradiated with different dose levels (0.5 and 5 Gy) in a 5 x 5 cm2 field at 6 MV. A previously developed solid phantom (polystyrene) was used with overall dimensions corresponding to an average human head. EBT films were placed at four different depths (10, 20, 25 and 30 mm) and all films were irradiated simultaneously. Four different treatment plans were verified that resemble typical clinical situations. These plans differed in irradiation technique (conformal mMLC or circular arc SRT) and in tumour size (PTV of 1 or 2.5 cm3). In-house developed software was applied to calculate gamma (gamma) index values and to perform several statistical operations (e.g. gamma-area histograms). At depths of 10 mm gamma1%, (gamma-value where 1% of the points have an equal or higher value in the region of interest) were between 1-3 and maximum gamma > 1 (% of gamma-values > 1 in the region of interest) areas were almost 30%. At larger depths, i.e. more close to the isocenter, gamma 1% was < 1 and gamma > 1 areas were mostly < 5%. Average gamma values were about 0.5. Besides the compromised accuracy in the buildup region, previously defined IMRT acceptance criteria [Stock et al., Phys. Med Biol. 50 (2005) 399-411] could be applied as well to SRT. Radiochromic EBT films, in combination with a flat-bed scanner, were found to be an ideal multidimensional dosimetric tool for treatment plan quality assurance. EBT films are a suitable and reliable dosimetric tool that could replace traditionally used radiographic films. The presented acceptance criteria for SRT treatment plans might be used as a benchmarking data-set for other stereotactic applications and/or other equipment (planning system and delivery hardware) combinations.
منابع مشابه
Evaluation of the RtDosePlan Treatment Planning System using Radiochromic Film and Monte Carlo Simulation
Introduction: GafChromic EBT films are one of the self-developing and modern films commercially available for dosimetric verification of treatment planning systems (TPSs). Their high spatial resolution, low energy dependence and near-tissue equivalence make them suitable for verification of dose distributions in radiation therapy. This study was designed to evaluate the dosimetric parameters of...
متن کاملComparison of EBT and EBT3 RadioChromic Film Usage in Parotid Cancer Radiotherapy
Background: EBT and EBT3 radioChromic films have been used in radiotherapy dosimetry for years.Objective: The aim of the current study is to compare EBT and EBT3 radioChromic films in dosimetry of radiotherapy fields for treatment of parotid cancer.Methods: Calibrations of EBT and EBT3 films were performed with identical setups using a 6 MV photon beam of a Siemens Primus linac. Skin dose was m...
متن کاملSkin Dosimetry in Radiotherapy of Breast Cancer: a Comparison between EBT and EBT3 Radiochromic Films
Objective: Radiochromic EBT3 film is a later generation of radiochromic films. The aim of this study is to compare EBT and EBT3 radiochromic films in radiotherapy fields of breast cancer.Methods: A RANDO phantom was irradiated by a 6 MV Siemens Primus linac with medial and lateral fields of radiotherapy of breast cancer. Dosimetry was performed in various points in the fields using EBT and EBT3...
متن کاملRadiation transmission, leakage and beam penumbra measurements of a micro‐multileaf collimator using GafChromic EBT film
Micro-multileaf collimator systems coupled to linear accelerators for radioneurosurgery treatments require a rigorous dosimetric characterization in order to be used in 3D conformal and intensity modulated stereotactic radiosurgery and radiotherapy applications. This characterization involves high precision measurements of leaf transmission, leakage and beam penumbra through the collimation sys...
متن کاملComparison of EBT and EBT3 RadioChromic Film Usage in Parotid Cancer Radiotherapy
BACKGROUND EBT and EBT3 radioChromic films have been used in radiotherapy dosimetry for years. OBJECTIVE The aim of the current study is to compare EBT and EBT3 radioChromic films in dosimetry of radiotherapy fields for treatment of parotid cancer. METHODS Calibrations of EBT and EBT3 films were performed with identical setups using a 6 MV photon beam of a Siemens Primus linac. Skin dose wa...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Zeitschrift fur medizinische Physik
دوره 18 1 شماره
صفحات -
تاریخ انتشار 2008