Technical performance evaluation of BM/Hitachi 747-200 serum fructosamine assay.

نویسندگان

  • M J Lin
  • C Hoke
  • B Ettinger
  • R V Coyne
چکیده

The technical performance of serum fructosamine assay as a short-term index of diabetic control was evaluated by using the BM/Hitachi 747-200 Automatic Analyzer (Boehringer Mannheim Corp). Intra- and interassay precisions (CV) were <2%. Linearity was confirmed up to 1000 micromol/L. Lipemia, produced artificially or from hypertriglyceridemia in vivo, did not interfere with measurement of serum fructosamine concentration. However, hemoglobin (Hb) interfered with measurement substantially at concentrations >15.5 micromol/L, and bilirubin interfered substantially at concentrations >68.4 micromol/L. A normal reference range of up to 282 micromol/L (mean +/- 2SD = 242 +/- 40) was derived from results obtained from 228 nondiabetic subjects. In 222 diabetic subjects, serum fructosamine concentration correlated well with fasting glucose concentration (r = 0.74) and with Hb A1c concentration (r = 0.80). This automated fructosamine assay has the advantages of technical simplicity, low cost, and reduced analytical time compared with the Hb A1c method.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of a random access analyser: BM/Hitachi 911

The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an ei...

متن کامل

Effects of various serum proteins on quantification of fructosamine.

We determined fructosamine concentrations with the CentrifiChem 600 centrifugal analyzer and the Hitachi 737 discrete analyzer. Reference intervals agreed with the most recently published results, and values in fasting patients were significantly correlated with glycated hemoglobin, plasma glucose, albumin, beta- and gamma-globulins, IgG, IgA, IgM, and total protein. Partial correlation analysi...

متن کامل

Relative potency of commercial calibrators for fructosamine, and their effect on measurements of fructosamine in serum.

The protein-based secondary standardsfor serum fructosamine in a commercial test kit (Roche Diagnostics) have recently been modified by the manufacturer. Under certain assay conditions, the new calibrator gave significantly greater serum fructosamine values than those obtained with the originally supplied calibrator. However, for 20-4, samples and absorbance measurements at 10 and 15 mm, hereaf...

متن کامل

Clinical validation of a second-generation fructosamine assay.

The serum fructosamine assay, used to monitor short-term clinical glycemic control, reportedly has several technical drawbacks. However, technical improvements have resulted in a new second-generation assay of fructosamine. We evaluated this second-generation assay (from Roche Diagnostics) in 529 nondiabetic and diabetic patients and found a highly significant correlation with results of the fi...

متن کامل

Fructosamine and glycated hemoglobin in the assessment of glycaemic control in dogs.

Fructosamine and glycated hemoglobin (HbA1c) concentrations were measured simultaneously in 222 dogs (96 healthy and 126 sick dogs). The dogs were divided into 3 groups according to the glucose concentration: hypo, hyper and euglycaemic dogs. Serum fructosamine concentrations were measured by the reduction test with nitroblue tetrazolium. A turbidimetric inhibition immunoassay and specific poly...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 42 2  شماره 

صفحات  -

تاریخ انتشار 1996