WE-G-217BCD-07: Implementation and Evaluation of Helical On-Board CBCT and Exact Image Reconstruction.

نویسندگان

  • J Tan
  • H Li
  • P Parikh
  • E Izaguirre
  • D Yang
چکیده

PURPOSE Longitudinal coverage of CBCT, which is 17 cm for head scan and 15.5 cm for body scan, is not enough to cover the entire PTV for over 90% of head/neck and gastrointestinal/genitourinary/gynecologic patients if lymph nodes are involved. Helical CBCT, which was accomplished using external beam LINAC in its research mode, is one promising way to extend the CBCT longitudinal coverage. Aim of this study is to compare Katsevich's exact algorithm with traditional FDK algorithm for helical CBCT image reconstruction. METHODS CBCT projection raw data were acquired on a TrueBeam LINAC machine (Varian Medical Systems) in the research mode in helical trajectories that encompass a 360 degree rotation, 25 cm pitch, 100 kVp, 80 mA, and 25 ms, with a Catphan 600. Reconstruction was done with Katsevich's exact and FDK approximate algorithms. Scatter correction, beam-hardening correction, and non-uniform gantry angle correction, are performed on projection data to reduce artifacts and noise. Image qualities (CT number accuracy, uniformity, SNR) were evaluated and compared between the reconstruction algorithms. RESULTS Images reconstructed by Katsevich's algorithm show better qualities, compared to ones by FDK algorithm and HU numbers have higher uniformity and accuracy. The HU-density calibration curve closely conforms to the manufacturer recommended values. The level of noise computed as the standard deviation in the phantom uniform region is 28.07 for the Katsevich algorithm, compared to 44.64 for the FDK algorithm. CONCLUSIONS Katsevich's exact reconstruction algorithm provided better image qualities than FDK for helical CBCT scans. This result will very useful for our ongoing investigation of helical CBCT, which would lead to improvement of CBCT longitudinal coverage of PTV and would be essential for future image-guided adaptive radiation therapy applications. Varian Research Agreement with Washington University in St. Louis.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Impact of Residual Geometric Inaccuracies on Normal Organ Doses in Image Guided-Radiation Therapy of Prostate Cancer Using On-Board Kilovoltage Cone-Beam Computed Tomography

Introduction: The aim of this retrospective study was to evaluate the variations in delivered dose to the bladder, rectum, and femoral heads of prostate cancer patients during a course of treatment by image-guided radiation therapy (IGRT). Materials and Methods: Overall, 15 patients with prostate cancer were selected and. Each week, for each patient five consecutive cone beam computed tomograph...

متن کامل

Methods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space

Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Materials and Methods: Compared to the traditional methods...

متن کامل

FPGA Implementation of JPEG and JPEG2000-Based Dynamic Partial Reconfiguration on SOC for Remote Sensing Satellite On-Board Processing

This paper presents the design procedure and implementation results of a proposed hardware which performs different satellite Image compressions using FPGA Xilinx board. First, the method is described and then VHDL code is written and synthesized by ISE software of Xilinx Company. The results show that it is easy and useful to design, develop and implement the hardware image compressor using ne...

متن کامل

Evaluation of the gray level in CBCT systems and its relationship with HU in CT Scanners

Introduction: Cone-beam CT (CBCT) is an imaging system which offers three-dimensional (3D), multiplanar images and has many advantages over computed tomography (CT) including shorter acquisition times for the resolution desired in dentistry, lower radiation dose to the patient, reasonable price and higher spatial resolution but CBCT scanners are unable to display actual Hounsf...

متن کامل

Evaluation of sparse-view reconstruction from flat-panel-detector cone-beam CT.

Flat-panel-detector x-ray cone-beam computed tomography (CBCT) is used in a rapidly increasing host of imaging applications, including image-guided surgery and radiotherapy. The purpose of the work is to investigate and evaluate image reconstruction from data collected at projection views significantly fewer than what is used in current CBCT imaging. Specifically, we carried out imaging experim...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 39 6Part28  شماره 

صفحات  -

تاریخ انتشار 2012