Inadvertent placement of pacemaker lead into the systemic ventricle in repaired D-transposition of the great arteries.
نویسندگان
چکیده
We report a case of a 27-year-old man with Senning repair of D-transposition of the great arteries. A dual-chamber pacemaker was placed for syncope due to sinoatrial node dysfunction 3 years before this presentation. He had no further syncope but complained of dyspnea on exertion and mild exercise intolerance. A routine chest x-ray and transthoracic echocardiography suggested the presence of a pacing lead in the systemic ventricle (anatomic right ventricle). To better define the anatomy and lead position, he underwent a gated computed tomography scan. This showed an appropriately positioned atrial lead in the nonsystemic atrium (anatomic left atrium), but the ventricular lead had perforated through the atrial baffle, coursing into the systemic ventricle (Figure, A and B; online-only Data Supplement Movies I and II). A small systemic-to-venous baffle shunt was seen in the region of the perforation. The systemic ventricle had mildly reduced systolic function. Inadvertent placement of pacing leads in the systemic ventricle and the presence of intracardiac shunts pose significantly high risk of systemic thromboembolism.1 During the 3 years since pacemaker implantation, our patient remained asymptomatic from thromboembolic complications. The lead was extracted from the systemic ventricle and reimplanted correctly in the nonsystemic ventricle (anatomic left ventricle) under fluoroscopic and transesophageal echocardiographic guidance (online-only Data Supplement Movies III and IV). Transthoracic echocardiography with bubble contrast after removal of lead from the systemic right ventricle showed a persistent baffle leak (online-only Data Supplement Movie V). This case highlights the importance of understanding the complex anatomy of the atrial baffles in patients with repaired transposition of the great arteries before device implantation to avoid potential complications from transvenous pacemaker implantation. It is also necessary to know the course of the venous baffle and rule out baffle leaks and obstruction. Newer imaging techniques such as cardiac computer tomography and magnetic resonance imaging are helpful in guiding management in complex congenital heart diseases. Compared with magnetic resonance imaging, computed tomography imaging is faster and can be used in patients with pacemakers and defibrillators.
منابع مشابه
Pacemaker laser lead extraction and reimplantation of dual-chamber implantable cardioverter defibrillator via Mustard baffle in complete transposition of great arteries.
We present a case of a complicated lead extraction and reimplantation of an implantable cardioverter defibrillator (ICD) in a young woman with complete transposition of great arteries (CTGA), a cyanotic congenital heart defect in which the aorta and the pulmonary trunk are transposed. The malformation results in two parallel circulations, whereby the left ventricle is attached to the pulmonary ...
متن کاملA 33-day-old Infant with the Transposition of the Great Arteries; A Rare Case Report
Dextro-transposition of the great arteries (d-TGA) is the one most common cyanotic congenital heart disease in neonates. The discordant ventriculoarterial arrangement results in parallel circulation, it so is vital to understand the management. We report a rare interesting but critical case of 33-day-old boy who developed cyanosis and had transposition of great arteries combined with interrupte...
متن کاملHemodynamic data and angiographic findings after Mustard repair for complete transposition of the great arteries.
SUMMARY Eleven childrern with uncomplicated transposition of the great arteries repaired by the Mustard procedure were studied by atrial pacing, cardiac catheterization, and ciiieangiography. Nine had persistent postoperative ectopic cardiac rhythms. Atrial pacing in five suggested sinus node damage in four and A-V junctional conduction tissue damage in one. In seven children dye-dilution curve...
متن کاملResynchronization therapy transvenous approach in dextrocardia and congenitally corrected transposition of great arteries.
Cardiac resynchronization therapy (CRT) is an acknowledged treatment for advanced heart failure in acquired dilated cardiomyopathy, resistant to pharmacotherapy. Although there are no therapeutic standards regarding heart failure originating from congenital heart defects with systemic right ventricle, a number of CRT implantations by transvenous approach in congenitally corrected transposition ...
متن کاملInadvertent malposition of a permanent pacemaker ventricular lead into the left ventricle which was initially missed and diagnosed two years later: a case report
INTRODUCTION Inadvertent malposition of a pacemaker ventricular lead into the left ventricle is an uncommon event, and its actual incidence is probably unknown. It may be underestimated and underreported because of a possible asymptomatic course. A 12-lead electrocardiogram is important to confirm proper placement. CASE PRESENTATION We report a case of a 60-year-old Caucasian man with a malpo...