Screening method to evaluate point-of-care human chorionic gonadotropin (hCG) devices for susceptibility to the hook effect by hCG β core fragment: evaluation of 11 devices.
نویسندگان
چکیده
BACKGROUND The predominant hCG variant in urine, hCG β core fragment (hCGβcf), has been demonstrated to cause false-negative results in qualitative point-of-care (POC) hCG devices. This is a major concern for healthcare professionals using POC pregnancy tests. We developed a screening method to evaluate qualitative POC hCG devices for their susceptibility to inhibition by hCGβcf. Using this method, we evaluated the performance of 11 commonly used devices. METHODS A wide range of purified hCG and hCGβcf concentrations were mixed and tested on 2 POC devices. By use of those results, a screening method was defined and 9 additional POC devices were evaluated. Two solutions containing (a) 500 pmol/L (171 IU/L) intact hCG with 0 pmol/L hCGβcf and (b) 500 pmol/L intact hCG with 500 000 pmol/L hCGβcf were used to screen all POC devices. RESULTS The OSOM and Cen-Med Elite devices were found to be most susceptible to false-negative results due to hCGβcf. The BC Icon 20 and the Alere were the least susceptible. The remaining 7 were moderately affected. Devices that gave the strongest signal with hCGβcf alone were those that were least likely to show a hook effect. CONCLUSIONS The screening method put forth here can be used by device users and manufacturers to evaluate POC devices for inhibition by hCGβcf. Of 11 devices evaluated, only 2 have been identified that exhibit minimal to no susceptibility to hCGβcf.
منابع مشابه
Qualitative point-of-care human chorionic gonadotropin testing: can we defuse this ticking time bomb?
Qualitative point-of-care (POC) human chorionic gonadotropin (hCG) testing is routinely performed on urine samples obtained in the Emergency Department (ED) or before surgery to rule out pregnancy and prevent fetal injury or death. Because potentially harmful treatments are allowed to proceed on the basis of a negative hCG test result, falsenegative results can have disastrous consequences. We ...
متن کاملEFFECT OF β-HUMAN CHORIONIC GONADOTROPIN (β-HCG) ON NEUTROPHIL FUNCTIONS
Many findings point to the presence of a close relationship between the immune, endocrine and neurologic systems. For example the suppressive effects of β-human chorionic gonadotropin (β-HCG) on IL-l production, the potentiative effect of enkephalins on lymphocyte functions and the suppressive effect of leuteinizing hormone (LH) on natural killer (NK) cell activities have been clearly esta...
متن کاملPoint-of-care and over-the-counter qualitative human chorionic gonadotropin (hCG) devices remain susceptible to false-negative results caused by excess hCG β core fragment.
Employment or Leadership: None declared. Consultant or Advisory Role: None declared. Stock Ownership: None declared. Honoraria: None declared. Research Funding: Bio-Rad Laboratories, ddPCR Reagents, Life Technologies, Acrometrix inhibition panel; K.R. Jerome, Bio-Rad Laboratories. Expert Testimony: None declared. Patents: None declared. Other Remuneration: T.C. Dingle, Bio-Rad Laboratories prov...
متن کاملThe Value of Serum Β-Subunit of Human Chorionic Gonadotropin Level in Prediction of Treatment Response to Methotrexate in Management of Ectopic Pregnancy; a Systematic Review and Meta-Analysis
Background: No consensus has been reached on prognostic value of serum concentration of β (beta) subunit of human chorionic gonadotropin (β-hCG) in treatment response to methotrexate in management of ectopic pregnancy. Therefore, the present study aimed to evaluate this subject through a systematic review and meta-analysis. Materials and Methods: An extensive literature search on online databas...
متن کاملSeparation of beta-human chorionic gonadotropin from fibrinogen using a MEMS size exclusion chromatography column
This article reports a MEMS (Micro-ElectroMechanical-Systems)-based column separator designed for potential integration with portable medical point-of-care testing (POCT) devices. The MEMS column uses size exclusion chromatography (SEC) to pre-separate raw samples by size, and is made of polydimethylsiloxane (PDMS) fabricated on a glass slide. The MEMS SEC column separates 300 ng/mL of beta-hum...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 60 4 شماره
صفحات -
تاریخ انتشار 2014