Analytical evaluation of Diazyme procalcitonin (PCT) latex-enhanced immunoturbidimetric assay on Beckman Coulter AU5800.
نویسندگان
چکیده
BACKGROUND This study was aimed to evaluate the analytical performance of the novel Diazyme procalcitonin (PCT) immunoturbidimetric assay on Beckman Coulter AU5800. METHODS Diazyme PCT is a latex-enhanced immunoturbidimetric assay, developed for use on laboratory instrumentations with capability of reading absorbance at 600 nm. This analytical evaluation included the assessment of limit of blank (LOB), limit of detection (LOD), functional sensitivity, imprecision, linearity, carryover, and method comparison between Diazyme PCT and Kryptor PCT on 129 routine serum inpatient samples. RESULTS The LOB, LOD, and functional sensitivity of Diazyme PCT were 0.16, 0.26, and 0.28 ng/mL, respectively. The intra- and inter-assay imprecision of Diazyme PCT was between 2.9% and 7.8%. The linearity was excellent in the range of PCT values between 0.16 and 56 ng/mL, and the carryover was negligible (0.02%). A highly significant agreement was found between Kryptor PCT and Diazyme PCT in a range of concentrations between 0.16 and 111 ng/mL (Diazyme PCT=1.10×Kryptor PCT-0.89; r=0.960; p<0.001). The mean bias was 0.48 ng/mL (95% CI, -0.58 to 1.54 ng/mL). The strength of agreement between Kryptor PCT and Diazyme PCT was between 85% and 96% at 0.50, 2.0, and 10 ng/mL cutoffs. CONCLUSIONS Diazyme PCT appears to be a reliable assay for diagnosis and management of critical care patients susceptible to severe bacterial infections.
منابع مشابه
Multicenter comparison of automated procalcitonin immunoassays
OBJECTIVES A multicenter study to compare results of BRAHMS Kryptor PCT with those obtained using four BRAHMS-partnered procalcitonin (PCT) automated immunoassays (DiaSorin Liaison, BioMérieux Vidas, Roche Cobas E601 and Siemens Advia Centaur) and the Diazyme immunotubidimetric assay implemented on four clinical chemistry platforms (Abbott Architect c16000, Siemens Advia 2400, Roche Cobas C501 ...
متن کاملPreliminary evaluation of the performance of a new, highly sensitive commercial immunoassay for serum ferritin determination.
We evaluated the analytical performance of a new, commercial, fully automated immunoturbidimetric assay for the determination of ferritin [FER-Latex(X2)CN SEIKEN, Denka Seiken, Japan] in serum on the Olympus AU2700 analyzer. The new assay is a latex-enhanced turbidimetric immunoassay with an analysis time of 10 min. The linearity of the assay was confirmed up to 2505 pmol/L (R2=0.999). The dete...
متن کاملAnalytical evaluation of the novel Lumipulse G BRAHMS procalcitonin immunoassay
OBJECTIVES This study was designed to evaluate the analytical performance of the novel Lumipulse G1200 BRAHMS procalcitonin (PCT) immunoassay. DESIGN AND METHODS This analytical evaluation encompassed the calculation of the limit of blank (LOB), limit of detection (LOD), functional sensitivity, intra- and inter-assay imprecision, confirmation of linearity and a comparison with the Vidas BRAHM...
متن کاملHyaluronic Acid: The biomarker for Liver Fibrosis An Automated Assay
The latex agglutination method by Wako (HA LT) shows excellent analytical characteristics that make it appropriate for clinical practice. The test is now easily automated and can be performed on most routine clinical chemistry analysers. For example, using a Beckman Coulter AU640 analyser Guéchot et al. measured the imprecision from within-run CV to be up to 2.12% and from between-run CV to be ...
متن کاملDetermining the concentration of procalcitonin using a magnetic particles-based chemiluminescence assay for the clinical diagnosis of sepsis.
Our objective is to develop an assay based on magnetic particles (MPs) to determine the concentration of procalcitonin (PCT) using a chemiluminescence immunoassay (CLIA). Fluorescein isothiocyanate (FITC) and N-(aminobutyl)-N-(ethylisoluminol) (ABEI) were used to label two different anti-procalcitonin (PCT) monoclonal antibodies. The labeled antibodies, the PCT antigen, and the anti-FITC antibo...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry and laboratory medicine
دوره 53 4 شماره
صفحات -
تاریخ انتشار 2015