An Analytical-empirical Calculation of Linear Attenuation Coefficient of Megavoltage Photon Beams

نویسندگان

  • F. Seif
  • M.J. Tahmasebi-Birgani
  • M.R. Bayatiani
چکیده

BACKGROUND In this study, a method for linear attenuation coefficient calculation was introduced. METHODS Linear attenuation coefficient was calculated with a new method that base on the physics of interaction of photon with matter, mathematical calculation and x-ray spectrum consideration. The calculation was done for Cerrobend as a common radiotherapy modifier and Mercury. RESULTS The values of calculated linear attenuation coefficient with this new method are in acceptable range. Also, the linear attenuation coefficient decreases slightly as the thickness of attenuating filter (Cerrobend or mercury) increased, so the procedure of linear attenuation coefficient variation is in agreement with other documents. The results showed that the attenuation ability of mercury was about 1.44 times more than Cerrobend. CONCLUSION The method that was introduced in this study for linear attenuation coefficient calculation is general enough to treat beam modifiers with any shape or material by using the same formalism; however, calculating was made only for mercury and Cerrobend attenuator. On the other hand, it seems that this method is suitable for high energy shields or protector designing.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An Analytical-empirical Calculation of Linear Attenuation Coefficient of Megavoltage Photon Beams

Background: In this study, a method for linear attenuation coefficient calculation was introduced.Methods: Linear attenuation coefficient was calculated with a new method that base on the physics of interaction of photon with matter, mathematical calculation and x-ray spectrum consideration. The calculation was done for Cerrobend as a common radiotherapy modifier and Mercury.Results: The values...

متن کامل

Monte Carlo calculation of shielding properties of newly developed heavy concretes for megavoltage photon beam spectra used in radiation therapy

Introduction: Globally, the need for radiotherapy as a part of cancer management increases every year. Thus, the shielding for megavoltage radiotherapy rooms is of great importance. Materials and Methods: In the present study, 14 types of developed high-density concrete with densities ranging from 2.45 to 5.11 were simulated by using Monte Carlo method. The linear attenuation coefficient and t...

متن کامل

Determination of the attenuation coefficient for megavoltage photons in the water phantom

Background: Attenuation coefficient (μ) plays an important role in calculations of treatment planning systems, as well as determination of dose distributions in external beam therapy, dosimetry, protection, phantom materials and industry. So, its exact measurement or calculation is very important. The aim of this study was to evaluate the μ in different points in the water phantom anal...

متن کامل

Investigation of the effects of a carbon-fiber tabletop on the surface dose and attenuation dose for megavoltage photon beams

Background: Multiple beams are generally used with an increased possibility that the beam axis intersects the treatment table. Treatment tabletops are commonly made of carbon fiber due to its high mechanical strength and rigidity, low specific density, extremely light and low radiation beam attenuation properties. Purpose of this paper is investigated the dose changes in the buildup region and ...

متن کامل

The evaluation of change in attenuation coefficient of cerrobend block used for radiation protection of healthy tissues in megavoltage photon radiation therapy after multiple melting

Introduction: Protecting the vital organs in radiation therapy is one of the most important issues. Cerrobend alloy, the most common material used as the shield, is melted several times and used for different patients. In this study changes in attenuation coefficient of cerrobend due to melting was investigated.   Materials and Methods: In melting furnace, ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2017