The influence of the level of experience of the laryngoscopist on the duration of laryngoscopy, the forces exerted on the tongue and on the maxillary incisors during laryngoscopy, were investigated. Five groups (anaesthetists, residents in anaesthesia, nurse anaesthetists, surgeons and medical students), each consisting of 15 individuals, participated in the study. An intubation manikin was used with a laryngoscope modified so that the forces applied during laryngoscopy could be measured. The mean duration of laryngoscopy in these groups was 23.4 sec, 17.6 sec, 27.1 sec, 26.8 sec and 42.7 sec, respectively. The maximally applied forces on the tongue were 71.7 N, 60.5 N, 65.9 N, 74.2 N and 69.7 N, respectively. The maximally applied forces on the maxillary incisors were 49.9 N, 36.3 N, 41.1 N, 58.3 N and 53.9 N, respectively. These results indicate the level of experience has a significant influence on the duration of laryngoscopy but seems to have little influence on the forces applied to the tongue and the maxillary incisors.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An evaluation of a modified Macintosh laryngoscope in a manikin.

PURPOSE In this manikin study a modified Macintosh blade was prospectively compared with its original focussing on the forces exerted on the maxillary incisor teeth and intubation success. The modified blade, a standard Macintosh blade with a reduced proximal flange, was intended to reduce the forces exerted on the maxillary incisors. METHODS A manikin equipped with two sensors, to measure fo...

متن کامل

Forces acting on the maxillary incisor teeth during laryngoscopy using the Macintosh laryngoscope.

We determined the forces on the maxillary incisors during routine laryngoscopy in 65 adult patients. The forces were measured by a strain gauge based sensor positioned between the handle and the blade of the laryngoscope. The mean maximal force acting on the maxillary incisors was 49 N. In patients without maxillary incisors, the force acting on the gums was significantly lower at 21 N (p < 0.0...

متن کامل

Reshaping the Macintosh blade using biomechanical modelling. A prospective comparative study in patients.

It has been demonstrated that during routine use of the Macintosh blade, great forces are exerted on the maxillary incisors. The aim of this study was, by using biomechanical modelling, to modify a standard Macintosh blade in order to reduce these forces. This resulted in a Macintosh blade with a reduced proximal flange. Five anaesthetists performed tracheal intubation in 46 patients using the ...

متن کامل

Evaluation of the success rate and quality of teaching of tracheal intubation to medical students by using video laryngoscope

Background: Training of airway management and tracheal intubation skill in emergencies for resuscitation of patients are basic skills for medical students. However, the success rate of beginners in this skill is low. Video laryngoscopes are new devices that can increase the success of endotracheal intubation training. Methods: This clinical trial was conducted with 30 medical students who came...

متن کامل

Conflicts of interest in expert-authored practice parameters, standards, guidelines, recommendations.

1. Amour J, Le Manach YL, Borel M, Lenfant F, Nicolas-Robin A, Carillion A, Ripart J, Riou B, Langeron O: Comparison of single-use and reusable metal laryngoscope blades for orotracheal intubation during rapid sequence induction of anesthesia: A multicenter cluster randomized study. ANESTHESIOLOGY 2010; 112:325–32 2. Amour J, Marmion F, Birenbaum A, Nicolas-Robin A, Coriat P, Riou B, Langeron O...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Canadian journal of anaesthesia = Journal canadien d'anesthesie

دوره 42 2  شماره 

صفحات  -

تاریخ انتشار 1995