Survey of the Performance of Commercial Dose Calibrators for Measurement of 123I Activity
نویسندگان
چکیده
The accuracy with which 123I activity is measured by a dose calibrator depends on the composition and geometry of the source. The present study assessed the variability of current commercial dose calibrators in assaying liquid 123I samples. Methods: A calibration procedure for 123I measurement was performed on 177 dose calibrators (11 manufacturers) at 138 sites in North America and Europe. Using the standard 123I push-button or dial setting, activity in a 5-mL 123I calibration source in a 10-mL U.S. Pharmacopeia type 1 glass vial (actual activity previously determined using a National Institute of Standards and Technology–traceable standard metrology chamber) was measured. A portion of the source was then transferred to a plastic syringe (10-, 5-, and 3-mL sizes at different sites), and the activity in the syringe was measured. Calibration factors (CFs) for converting the dose calibrator readings to the reference activities were then determined for the vial and the syringes. Data were analyzed for all calibrators combined and based on device manufacturers. Measurements using a copper attenuator (sleeve) were made for a subset of 10 dose calibrators at sites that used these devices in clinical practice. Results: Mean CFs for the different measurements were as follows: 10-mL vial, 1.278; 10-mL syringe, 0.811; 5-mL syringe, 0.815; 3-mL syringe, 0.792. Almost half of the dose calibrators had vial CFs between 1.2 and 1.3 and 10-mL syringe CFs between 0.7 and 0.8, whereas less than 16% of the instruments had uncorrected readings within 610% of the reference activities. Although there was a wide range of CFs for different calibrators using the copper sleeve, for each unit the CFs for the glass vial and the 10-mL plastic syringe were virtually identical. Conclusion: On most commercial dose calibrators, the standard 123I settings result in significant errors in activity measurements for sources in glass vials and plastic syringes. The difference in ionization chamber detection caused by the container composition (glass vs. plastic) is a much larger source of measurement variation than source volume or geometry.
منابع مشابه
An investigation on the performance of dose calibrators in nuclear medicine centers in Iran
Background: To investigate the status of the nuclear medicine (NM) centers in Iran for the performance of dose calibrators, 18 out of 54 centers providing NM services in Iran were randomly selected and inspected in 1997. In the first phase of the study the selected centers were inspected for performing of quality control (QC) tests of dose calibrators. The linearity of the activity res...
متن کاملSurvey of the performance of commercial dose calibrators for measurement of ¹²³I activity.
UNLABELLED The accuracy with which (123)I activity is measured by a dose calibrator depends on the composition and geometry of the source. The present study assessed the variability of current commercial dose calibrators in assaying liquid (123)I samples. METHODS A calibration procedure for (123)I measurement was performed on 177 dose calibrators (11 manufacturers) at 138 sites in North Ameri...
متن کاملUptake index of 123I-metaiodobenzylguanidine myocardial scintigraphy for diagnosing Lewy body disease
Objective(s): Iodine-123 metaiodobenzylguanidine (123I-MIBG) myocardial scintigraphy has been used to evaluate cardiac sympathetic denervation in Lewy body disease (LBD), including Parkinson’s disease (PD) and dementia with Lewy bodies (DLB). The heart-tomediastinum ratio (H/M) in PD and DLB is significantly lower than that in Parkinson’s plus syndromes and Alzheimer’s disease. Although this ra...
متن کاملDevelopment of an 123I-metaiodobenzylguanidine Myocardial Three-Dimensional Quantification Method for the Diagnosis of Lewy Body Disease
Objective(s): We recently developed a new uptake index method for 123I-metaiodobenzylguanidine (123I-MIBG) heart uptake measurements by using planar images (radioisotope angiography and planar image) for the diagnosis of Lewy body disease (LBD), including Parkinson’s disease (PD) and dementia with Lewy bodies(DLB). However, the diagnostic accuracy of the uptake index was approximately equal to ...
متن کاملMeasurement of the Radiation Dose Rates of Patients Receiving Treatment with I-131 Using Telescopic Radiation Survey Meter
Introduction In order to discharge the patients receiving treatment with large radiation doses of 131I for thyroid cancer, it is necessary to measure and evaluate the external dose rates of these patients. The aim of the study was to assess a new method of external dose rate measurement, and to analyze the obtained results as a function of time. Materials and Methods In this study, a telescopic...
متن کامل