Early Malfunction of Transvenous Pacemaker Electrodes

نویسندگان

  • ELIZABETH CASSADY
  • Robyn Lurie
چکیده

A 3-year study by three medical centers has revealed a 1-year electrode malfunction rate of 7.4%; most malfunctions occurred within the first 30 days. The incidence of unavoidable early malfunction (3.2%) fell within the 5% standards suggested by the committee report of the Inter-Society Committee on Heart Diseases. Incidences of obscure cause (3.2%) may be difficult to identify prospectively and may be, to a certain extent, unavoidable. The majority of the malfunctions (4.2%) showed specific clues that indicated that they were preventable. Successful repositioning was achieved on the first attempt in 80.6% of the cases with malfunction, and only 0.7% required ultimate myocardial electrode implantation. The principal clues to potentially unsatisfactory positioning included the presence of a large right ventricle with or without tricuspid insufficiency, current thresholds greater than 0.5 mA and ST-segment deviations on the intracardiac electrogram of less than 2 mV. Electrode malfunction may be more common with bipolar than with unipolar electrodes; but significant differences in the incidence of malfunction among different unipolar electrodes were observed. These data indicate that further developments in transvenous electrode design are warranted.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

ffect of Chlorhexidine on Prevention of Local Infection in Temporary Transvenous Pacemaker Catheter Site

Background and Objectives: The local infection of temporary transvenous pacemaker catheter site is a common hospital acquired infections. This study aimed to assess the effect of the Chlorhexidine Glaciate 2% on prevention of local infection on temporary transvenous pacemaker catheter site. Materials and Methods: In this 7-month randomized controlled trial, 100 patients having temporary transve...

متن کامل

Device for implantation of long-term transvenous cardiac pacemaker electrodes.

General anaesthesia is usually recommended for the most painful part of long-term transvenous pacemaker implantation, viz., the pulling subcutaneously of the end of the electrode from the neck incision to the site of pacemaker implantaition. The ability to perform this step, and therefore the whole procedure, under local anaesthesia is an advantage. We descrilbe a simple, inexpensive device whi...

متن کامل

Clinical evaluation of myocardial perforation as a complication of permanent transvenous pacemakers.

A simple and reliable radiographic method to aid in the detection of penetration or perforation of the myocardium by permanent transvenous pacemaker electrodes is described. The "epicardial fat-pad sign" wm M y positive in three Usymptomatic" patients in whom electrodes had perforated the myocardium and in eight of 28 "nonsymptomatic" patients with pacemakers implanted ra long arr three years. ...

متن کامل

Permanent ventricular pacemakers. Comparison of transthoracic and transvenous implantation.

In the series of 86 patients with the Stokes-Adams syndrome or with symptomatic bradycardia managed with permanently implanted ventricular pacemakers reported on, 35 patients had primary implantation of epicardial leads at thoracotomy and 51 had transvenous endocardial electrodes passed via the jugular venous system for permanent ventricular pacing. The transvenous method of permanent pacemaker...

متن کامل

Permanently Implanted Transvenous

Three patients with complete heart block had a high resistance to electrical stimulation of the endocardium through transvenous pacemaker systems. One patient died; one needed reoperation for implantation of epicardial electrodes; and the third required a generator of higher current. The present enthusiasm for permanent transvenous implantation of pacemakers must be tempered somewhat by the kno...