CARDIOVASCULAR MEDICINE Effect of atrial antitachycardia pacing treatments in patients with an atrial defibrillator: randomised study comparing subthreshold and nominal pacing outputs

نویسنده

  • A R J Mitchell
چکیده

Objective: To assess the true efficacy of antitachycardia pacing on spontaneous persistent atrial fibrillation in patients with an implanted atrial defibrillator, by comparing the effects of nominal pacing treatment with subthreshold pacing treatment. Design: The effects of antitachycardia pacing and burst 50 Hz atrial pacing on spontaneous atrial arrhythmias were evaluated six months after implantation of a Medtronic Jewel AF atrial defibrillator. Setting: Cardiology department in a district general hospital. Patients: 15 patients with persistent atrial fibrillation. Interventions: Patients were randomised to either “nominal” output pacing treatment or surface ECG and endocardial electrogram proven subthreshold “sham” pacing treatment for three months, and then crossed over to the alternative treatment for a further three months. Results: During the nominal output phase, 31 episodes of atrial fibrillation were treated with 53 bursts of 50 Hz pacing, 98 sequences of ramp atrial pacing, and 61 sequences of burst atrial pacing. Atrial fibrillation was not pace terminated during any episode. Thirty one episodes of atrial tachycardia were treated with 19 bursts of 50 Hz atrial pacing, 103 sequences of ramp atrial pacing, and 38 sequences of burst atrial pacing. Termination of atrial tachycardia was observed in 17 episodes. During the “sham” pacing period, no episodes were terminated by any pacing treatment. Conclusion: Atrial antitachycardia pacing treatments are ineffective at terminating persistent atrial fibrillation but may be useful in terminating episodes of atrial tachycardia or flutter, thus reducing the burden of arrhythmia.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of atrial antitachycardia pacing treatments in patients with an atrial defibrillator: randomised study comparing subthreshold and nominal pacing outputs.

OBJECTIVE To assess the true efficacy of antitachycardia pacing on spontaneous persistent atrial fibrillation in patients with an implanted atrial defibrillator, by comparing the effects of nominal pacing treatment with subthreshold pacing treatment. DESIGN The effects of antitachycardia pacing and burst 50 Hz atrial pacing on spontaneous atrial arrhythmias were evaluated six months after imp...

متن کامل

self-adapting autodecremental overdrive pacing

The efficacy and safety of a new antitachycardia pacing technique, self-adapting decremental overdrive pacing, was assessed in patients with clinical ventricular tachyarrhythmias who underwent programmed ventricular stimulation and serial drug testing. The three phases of this study involved a learning/experience phase, followed by intrapatient comparison of decremental overdrive pacing with co...

متن کامل

Catheter ablation versus standard conventional treatment in patients with left ventricular dysfunction and atrial fibrillation (CASTLE-AF) - study design.

BACKGROUND Electrical isolation of the pulmonary veins by catheter ablation is an emerging treatment modality for the treatment of atrial fibrillation (AF) and is increasingly used in patients with heart failure. METHODS The catheter ablation versus standard conventional treatment in patients with left ventricular dysfunction and atrial fibrillation trial (CASTLE-AF) is a randomized evaluatio...

متن کامل

Detection of Atrial Fibrillation and Flutter by a Dual-Chamber Implantable Cardioverter-Defibrillator

Background—To distinguish prolonged episodes of atrial fibrillation (AF) that require cardioversion from self-terminating episodes that do not, an atrial implantable cardioverter-defibrillator (ICD) must be able to detect AF continuously for extended periods. The ICD should discriminate between atrial tachycardia/flutter (AT), which may be terminated by antitachycardia pacing, and AF, which req...

متن کامل

Dual chamber pacing mode in an atrial antitachycardia pacing device without a ventricular lead - A necessary evil.

We present a case of a single chamber atrial pacemaker implanted for sinus node dysfunction and treatment of macroreentrant atrial tachycardias with atrial antitachycardia pacing. The patient presented with sustained atrial tachycardia above the detection rate, however, the device was unable to detect the tachycardia and did not deliver the programmed therapy. We discuss the nuances of the atri...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2002