Earfold Implantable Clip System for Correction of Prominent Ears: Analysis of Safety in 403 Patients
نویسندگان
چکیده
Background The Earfold system, a new treatment for the correction of prominent ears, consists of 3 components: the Earfold implant, the Earfold introducer, and the Prefold positioner. Methods This is an interim report based on an ongoing analysis of safety in a series of patients treated for prominent ears with the Earfold implant between February 2013 and September 2014. Safety was assessed based on adverse event reports and the need for implant revision; follow-up is ongoing. Results Seven surgeons used 1,200 Earfold implants to treat 403 patients (ages, 7-70 years; 63% male); the time since the initial implant procedure now ranges from 30 to 48 months. To date, 145 patients (36%) have returned for a follow-up visit (mean, 7.7 months [range, 1-34 months]). Adverse events requiring intervention have affected 39 of 403 (9.7%) patients; these include implant revisions (n = 17 [4.2%], most often due to implant visibility), skin erosion over the implant (n = 15 [3.7%]), and infection (n = 7 [1.7%]). Bleeding, recurrence of prominence, hematoma, deformity, or adverse scarring did not occur. Conclusions This interim analysis has shown that Earfold prominent ear correction system is associated with relatively few adverse events that require intervention; a small number of patients experienced infection, implant extrusion, or implant visibility that required revision. Most adverse events were related to either patient selection or technical errors at implantation. It is expected that with continued use of Earfold by surgeons experienced in otoplasty, the adverse event incidence will decrease.
منابع مشابه
Treatment of Prominent Ears with an Implantable Clip System: A Pilot Study.
BACKGROUND The earFold™ implantable clip system is a new treatment for prominent ears using an implant made from nickel-titanium alloy, forged into a predetermined shape. The implant is fixed to the cartilage then released, causing the cartilage to fold back. OBJECTIVES The study aimed to test the safety and behaviour of the implant in vivo. METHODS This was a Phase 1, prospective, nonrando...
متن کاملComparison of Fluoroplastic Causse Loop Piston and Titanium Soft-Clip in Stapedotomy
Introduction:Different types of prosthesis are available for stapes replacement. Because there has been no published report on the efficacy of the titanium soft-clip vs the fluoroplastic Causse loop Teflon piston, we compared short-term hearing results of both types of prosthesis in patients who underwent stapedotomy due to otosclerosis.Materials and Methods:A total of 57 ears were included in ...
متن کاملIs correction for metallic artefacts mandatory in cardiac SPECT/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?
Introduction: Metallic artifacts due to pacemaker/ implantable cardioverter defibrillator (ICD) leads in CT images can produce artifactual uptake in cardiac SPECT/CT images. The aim of this study was to determine the influence of the metallic artifacts due to pacemaker and ICD leads on myocardial SPECT/CT imaging. Methods: The study included 9 patients wh...
متن کاملEar molding in newborn infants with auricular deformities.
BACKGROUND A review of a single physician's experience in managing over 831 infant ear deformities (488 patients) is presented. METHODS The authors' methods of molding have advanced from the use of various tapes, glues, and stents, to a comprehensive yet simple system that shapes the antihelix, the triangular fossa, the helical rim, and the overly prominent conchal-mastoid angle (EarWell Infa...
متن کاملدستگاههای کاشتپذیر برای کودکان ناشنوا
The idea of providing hearing to profoundly deaf patients by artificially stimulating the sensory system has progressed from a futuristic possibility to reality. Generally, an implantable hearing device is designed to capture sound and present it to the auditory system, where at least part of the device is surgically implanted in the child receiving it. The goal of the implantable devices in yo...
متن کامل