Comparison of measured and computed portal dose for IMRT treatment

نویسندگان

  • Savino Cilla
  • Pietro Viola
  • Luigi Azario
  • Luca Grimaldi
  • Maurizio Craus
  • Guido D'Onofrio
  • Andrea Fidanzio
  • Francesco Deodato
  • Gabriella Macchia
  • Cinzia Digesù
  • Alessio G. Morganti
  • Angelo Piermattei
چکیده

A new 2D array Seven 29T model (PTW, Freiburg), equipped with 729 vented plane-parallel ion chambers, projected for pretreatment verification of radiotherapy plans, was used as a detector for the transmitted or portal dose measurements below a Rando phantom. The dosimetric qualities of the 2D array make it attractive for measuring transmitted dose maps from step-and-shoot intensity-modulated radiotherapy (IMRT). It is well known that for step-and-shoot IMRT beams that use a small number of monitor units (MUs) per sequence, the early and recent electronic portal imaging devices (EPIDs) present a different response at X-ray start-up that affects the accuracy of the measured transmitted dose. The comparison of portal doses measured to those calculated by a commercial treatment-planning system (TPS) can verify correct dose delivery during treatment. This direct validation was tested by irradiating a simulated head tumor in a Rando anthropomorphic phantom by step-and-shoot IMRT beams. The absolute transmitted doses on a plane orthogonal to the beam central axis below the phantom were measured by the 2D array calibrated in terms of dose to water and compared with the computed portal dose extracted by custom software. In a previous paper, the comparison between the IMRT portal doses, computed by a commercial TPS and measured by a linear array that supplied a 1 mm spatial dose resolution, was carried out. The gamma-index analysis supplied an agreement of more than 95% of the dose point with acceptance criteria, in terms of dose difference, DeltaDmax, and distance agreement, deltadmax, equal to 4% and 4 mm, respectively. In this paper, we verify the possible use of the PTW 2D array for measurements of the transmitted doses during several fractions of head and neck tumor radiotherapy. There are two advantages in the use of this 2D array as a portal dose device for the IMRT quality assurance program: first is the ability to perform absolute dose comparisons for hundreds of measurement positions to verify the correct dose delivery in several fractions of the therapy; second is the efficiency in time to detect these kinds of dose distributions within the field of view area of the CT scanner.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

متن کامل

Dosimetric comparison of 3-dimensional conformal and intensity-modulated radiotherapy techniques for whole breast irradiation in the prone and supine positions

Background: The aim of this study was to compare the differences of the dosimetric parameters between three-dimensional conformal radiotherapy (3D-CRT) and simultaneous-integrated boost intensity-modulated radiotherapy (SIB-IMRT) techniques in the prone and supine positions for breast irradiation. Materials and Methods: Ten patients underwent a computed tomography simulation in both the prone a...

متن کامل

Large discrepancies between planned and actually delivered dose in IMRT of head and neck cancer. A case report.

The case is reported of a patient with locally recurrent carcinoma of the tongue treated with intensity-modulated radiotherapy (IMRT) (simultaneous integrated boost) plus concurrent chemotherapy, who during the third week of radiotherapy developed grade 3 mucositis. Treatment was interrupted for 10 days until significant resolution of the symptoms. At the time of treatment resumption the patien...

متن کامل

Radiation treatment planning for prostate cancer: A new dosimetric comparison of five and seven fields IMRT plans

Background: To compare the dosimetric coverage of the planning target volume (PTV) and the dose delivered to the main Organs at Risk (OARs) in 5 and 7-field techniques of Intensity Modulated Radiation Therapy (IMRT) in patients with local prostate cancer. Materials and Methods: Twelve patients with local prostate cancer underwent 5 and 7-field IMRT planning. The delivery of IMRT was carried out...

متن کامل

Dose Distribution Analysis of Rapid Arc and Intensity Modulated Radiotherapy Plan in Head and Neck Cancer

Introduction: Globally, intensity-modulated radiation therapy (IMRT) is considered as highly precise and accurate method of radiotherapy planning. This technique amplifies spatial dose distribution conformity by modulating the intensity of radiation beams in each sub-volume. Additionally, it can reduces the dose to surrounding critical organs and deliver the planned dose to targets with the nom...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2006