Photon Attenuation Correction in Misregistered Cardiac PET/CT

نویسندگان

  • Axel Martinez-Möller
  • Nassir Navab
  • Markus Schwaiger
  • Stephan G. Nekolla
چکیده

PET-CT misalignment has been reported as a source of artefacts in cardiac PET/CT due to a biased photon attenuation correction. Using the data from 28 cardiac PET/CT rest/stress examinations, PETCT misalignment was corrected using three different methods: manual registration, automatic mutual information based image registration and an emission-driven correction algorithm. The clinical effects of the realignment were quantitatively assessed, and significant changes were observed in 6 out of 28 examinations (21.4%), in 5 of these studies resulting in the disappearance of large apparent perfusion defects. These results indicate that the excellent specificity of PET for the detection of perfusion defects could be compromised when CT-based attenuation correction is done without correction of PET-CT misalignment.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions.

UNLABELLED CT-based attenuation correction is a widely used option in commercial PET/CT scanners. However, as a result of a nonsimultaneous acquisition and differences in temporal resolution between both modalities, a potential misregistration between the PET and CT, especially in the thorax and the upper abdomen, can be found. We observed a substantial number of apparent perfusion defects in s...

متن کامل

The influence of misregistration between CT and SPECT images on the accuracy of CT-based attenuation correction of cardiac SPECT/CT imaging: Phantom and clinical studies

Introduction: Integration of single photon emission computed tomography (SPECT) and computed tomography (CT) scanners into SPECT/CT hybrid systems permit detection of coronary artery disease in myocardial perfusion imaging (MPI). Misregistration between CT and emission data can produce some errors in uptake value of SPECT images. The aim of this study was evaluate the influence...

متن کامل

Computed tomography based attenuation correction in PET/CT: Principles, instrumentation, protocols, artifacts and future trends

  The advent of dual-modality PET/CT imaging has revolutionized the practice of clinical oncology, cardiology and neurology by improving lesions localization and the possibility of accurate quantitative analysis. In addition, the use of CT images for CT-based attenuation correction (CTAC) allows to decrease the overall scanning time and to create a noise-free attenuat...

متن کامل

Cine CT for attenuation correction in cardiac PET/CT.

UNLABELLED In dual-modality PET/CT systems, the CT scan provides the attenuation map for PET attenuation correction. The current clinical practice of obtaining a single helical CT scan provides only a snapshot of the respiratory cycle, whereas PET occurs over multiple respiratory cycles. Misalignment of the attenuation map and emission image because of respiratory motion causes errors in the at...

متن کامل

A novel dual energy CT-based attenuation correction method in PET/CT systems: A phantom study

  In present PET/CT scanners, PET attenuation correction is performed by relying on the information given by CT scan. In the CT-based attenuation correction methods, dual-energy technique (DECT) is the most accurate approach, which has been limited due to the increasing patient dose. In this feasibility study, we have introduced a new method that can implement dual-en...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2007