Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements.

نویسندگان

  • Edward Ebramzadeh
  • Sophia N Sangiorgio
  • Federico Lattuada
  • Joon-Soon Kang
  • Roberto Chiesa
  • Harry A McKellop
  • Lawrence D Dorr
چکیده

BACKGROUND Although a number of methods are used to estimate polyethylene liner wear from radiographs of total hip replacements, there is no consensus with regard to the accuracy of these methods. The purpose of this study was to compare the accuracy of several such measurement methods with use of both laboratory radiographs and routine clinical radiographs. METHODS A phantom apparatus was designed to simulate random values of three-dimensional wear, with varying degrees of cup abduction and anteversion, and to obtain anteroposterior and cross-table lateral radiographs with each value. Wear was measured with use of the Charnley duoradiographic method, the Livermore method, and the method described by Dorr and Wan, as well as with use of PolyWare and Hip32 software packages, both with and without three-dimensional measurements. Clinical wear was measured from conventional radiographs made prior to revision surgery in fourteen patients and was compared with wear measured directly from the retrieved liners with use of a coordinate measuring machine. RESULTS With laboratory radiographs, median errors were 0.1 mm with the Livermore method and both computerized methods, 0.23 mm with the Charnley method, and 1.7 mm with the method of Dorr and Wan. Maximum errors were between 0.6 mm (Livermore) and 4.3 mm (Dorr and Wan). In contrast, with use of clinical radiographs, median errors ranged between 0.2 mm (Hip32) and 0.6 mm (Dorr and Wan). Maximum errors ranged between 1.8 mm (Dorr and Wan) and 2.5 mm (Livermore). CONCLUSIONS With laboratory radiographs, computerized methods of polyethylene wear measurement offered distinctly greater accuracy than did manual methods; however, with clinical radiographs, they offered only slightly better accuracy. Although the increased accuracy of computerized methods may be necessary in research settings, manual methods provided sufficient accuracy for routine clinical assessment of wear.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Automated assessment of polyethylene wear in cemented acetabular components using anteroposterior radiographs of total hip replacements

Polyethylene wear in the acetabular components of hip prostheses is implicated in loosening and failure. Radiographic measurement of wear is used to identify patients at risk and to assess prosthesis designs. This paper focuses on analysis of prostheses with cemented acetabular cups from anteroposterior (AP) radiographs. The articular surface of the femoral head and the acetabular rim marker ar...

متن کامل

Investigation of Wear Behavior of Biopolymers for Total Knee Replacements Through Invitro Experimentation

The average life span of knee prosthesis used in Total Knee Replacement (TKR) is approximately 10 to 15 years. Literature indicates that the reasons for implant failures include wear, infection, instability, and stiffness. However, the majority of failures are due to wear and tear of the prosthesis. The most common biopolymer used in TKR  is Ultra High Molecular Weight Polyethylene (UHMWPE). Pr...

متن کامل

Observations on retrieved humeral polyethylene components from reverse total shoulder arthroplasty.

HYPOTHESIS Analyses of polyethylene components retrieved at revision of total knee, hip, and shoulder replacements have been used to study the effect of design, patient, and surgical factors on initial implant performance, but no studies have reported similar types of findings in retrieved humeral polyethylene components in reverse total shoulder arthroplasty. Our hypothesis is that while the c...

متن کامل

Analysis of polyethylene wear in plain radiographs

BACKGROUND AND PURPOSE Two-dimensional computerized radiographic techniques are frequently used to measure in vivo polyethylene (PE) wear after total hip arthroplasty (THA), and several variables in the clinical set-up may influence the amount of wear that is measured. We compared the repeatability and concurrent validity of linear PE wear on plain radiographs using the same software but a diff...

متن کامل

More than 50% reduction of wear in polyethylene liners with alumina heads compared to cobalt-chrome heads in hip replacements

BACKGROUND AND PURPOSE Excessive wear of acetabular liners in hip replacements may lead to osteolysis and cup loosening. Different head materials are currently used. We measured differences in wear between alumina and cobalt-chrome heads with the same polyethylene liner. PATIENTS AND METHODS 39 patients (43 hips) with osteoarthritis were included in a study with 10-year follow-up. Wear was me...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. American volume

دوره 85-A 12  شماره 

صفحات  -

تاریخ انتشار 2003