Inaccuracies in using aortic valve gradients alone to grade severity of aortic stenosis.

نویسندگان

  • M J Griffith
  • C Carey
  • D J Coltart
  • B S Jenkins
  • M M Webb-Peploe
چکیده

The severity of aortic stenosis is an important determinant of prognosis in patients with symptoms who do not undergo valve replacement. To assess the pitfalls of using valve gradients alone 636 patients with aortic stenosis in whom the aortic valve area had been calculated by the Gorlin formula were studied. The correlation between valve area and aortic gradients was poor. No gradient was found that was both sensitive and specific for aortic stenosis. The maximum predictive accuracy was 81% for a mean gradient of 30 mm Hg and 80% for a peak gradient of 30 mm Hg. A mean gradient of 50 mm Hg or a peak gradient of 60 mm Hg were specific with a 90% or more positive predictive value. It proved difficult, however, to find a lower limit with a 90% negative predictive value. Patients with severe aortic stenosis and low gradients (peak or mean gradient of less than 30 mm Hg) had small ventricles (on both angiographic and echocardiographic data) with good ejection fractions and so were unlikely to be detected subjectively. In comparison patients with mild aortic stenosis and low gradients tended to have more aortic regurgitation but have similar degrees of left ventricular hypertrophy on echocardiographic or electrocardiographic criteria. The aortic valve area should be measured in all patients with the suspicion of severe aortic stenosis with a mean gradient of less than 50 mm Hg (50% of patients in this study) or a peak gradient of less than 60 mm Hg (47% of patients in this study).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

بررسی نتایج پیگیری یک ساله بالون والولوپلاستی تنگی مادرزادی دریچه آئورت در کودکان به مدت 10 سال در بیمارستان قلب شهید رجایی

Abstract Background: Balloon valvuloplasty uses for treatment of aortic stenosis from 20 years ago. Regarding loss of information about results of balloon valvuloplasty of hereditary aortic stenosis, this study was done to determine prognosis and complication of balloon valvuloplasty during last 10 years. Methods: This study was done on existing data of all children with hereditary aortic...

متن کامل

Bicuspid Aortic Valve: An Unusual Cause of Aneurysm of Left Coronary Sinus of Valsalva

Bicuspid aortic valve is traditionally considered an innocuous congenital anomaly. Due to a better and widespread availability of non-invasive imaging techniques, it has come to the fore that 30% of these cases develop complications, viz., valve abnormality (aortic regurgitation and stenosis), and aneurysm of aortic root and ascending aorta. Sinus of Valsalva aneurysm is an uncommon complicatio...

متن کامل

مقایسه نتایج درمانی بالون والوپلاستی و والوتومی جراحی در بیماران مبتلا به تنگی مادرزادی دریچه آئورت در پیگیری 6 ساله در بخش کودکان بیمارستان قلب شهید رجایی تهران

Background: Balloon valvuloplasty is a relatively new technique for relief of obstruction of aortic valve. The primary results of this procedure are similar to surgical valvotomy with fewer complications and mortality. Materials and methods: One hundred fifty-five patients with congenital aortic stenosis were enrolled in this study over a period of 6 years (1991-1997). Participants were all un...

متن کامل

Limitations in assessing the severity of aortic stenosis by Doppler gradients.

Continuous wave Doppler echocardiography was performed before cardiac catheterisation in 69 consecutive patients with suspected aortic stenosis. Agreement between the maximum and the mean Doppler gradients and catheterisation gradients was good. Doppler echocardiography, however, systematically underestimated the maximum and mean gradients, particularly in the high range. Stepwise regression an...

متن کامل

Transcatheter Aortic Valve Replacement: A Review Article

Transcatheter aortic valve replacement (TAVR) is a novel therapeutic intervention for the replacement of severely stenotic aortic valves in high-risk patients for standard surgical procedures. Since the initial PARTNER trial results, use of TAVR has been on the rise each year. New delivery methods and different valves have been developed and modified in order to promote the minimally invasive p...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • British heart journal

دوره 62 5  شماره 

صفحات  -

تاریخ انتشار 1989