The Combat Ready Clamp (CRoCTM) was designed to control hemorrhage from the groin region, on the battlefield. The purpose of this experiment was to determine whether CRoCTM user performance varied by the surface the casualty laid on (flat-hard, flat-soft, and curved-soft) and how quickly the device could be applied. The commercial manikin selected to assess user performance was designed to train soldiers in CRoCTM use. The manikin simulated severe hemorrhage from an inguinal wound, controllable by correct use of the CRoCTM. Each individual (n = 6) performed 3 iterations on each of the 3 surfaces (54 iterations total). The CRoCTM achieved hemorrhage control 100% of the time (54/54). Patient surface affected time to stop bleeding. The flat-soft surface (padded, 55 ± 9.7 seconds) was significantly different from the curved-soft surface (litter, 65 ± 16.5 seconds) and had the lowest overall total time (p = 0.007); time for the hard-flat surface was 58 ± 9.5 seconds. Users were trained to use the Combat Ready Clamp effectively, and the surface the casualty was lying on made some difference to user performance. All six persons trained had success in all nine of their iterations of CRoCTM use– a 100% rate. These findings indicate that training was effective and that training of other users is plausible, feasible, and practical within the scope of the present evidence.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of users to control simulated junctional hemorrhage with the combat ready clamp (CRoC™).

The Combat Ready Clamp (CRoC™) was designed to control hemorrhage from the groin region, on the battlefield. The purpose of this experiment was to determine whether CRoC™ user performance varied by the surface the casualty laid on (flat-hard, flat-soft, and curved-soft) and how quickly the device could be applied. The commercial manikin selected to assess user performance was designed to train ...

متن کامل

Testing of junctional tourniquets by military medics to control simulated groin hemorrhage.

BACKGROUND Junctional hemorrhage is a common cause of death on the battlefield, but there is no documented direct comparison for the use of junctional tourniquet models by US medics. The purpose of this testing is to assess military medic experience with the use of junctional tourniquets in simulated out-of-hospital trauma care. METHODS Nine medics (seven men and two women) used four differen...

متن کامل

In vivo assessment of the Combat Ready Clamp to control junctional hemorrhage in swine.

BACKGROUND Junctional wounds and associated hemorrhage have become more common and more lethal in the current war. The Combat Ready Clamp (CRoC) has been developed and deployed for treating junctional hemorrhage on the battlefield. This study examined the efficacy of CRoC and its acute effects in an animal model. METHODS Anesthetized pigs (n = 6) were subjected to laparotomy, splenectomy, and...

متن کامل

First case report of SAM(r) Junctional tourniquet use in Afghanistan to control inguinal hemorrhage on the battlefield.

Junctional hemorrhage, bleeding that occurs at the junction of the trunk and its appendages, is the most common preventable cause of death from compressible hemorrhage on the battlefield. As of January 2014, four types of junctional tourniquets have been developed and cleared by the U.S. Food and Drug Administration (FDA). Successful use of the Abdominal Aortic Tourniquet (AAT™) and Combat Read...

متن کامل

Prehospital emergency inguinal clamp controls hemorrhage in cadaver model.

BACKGROUND The Combat Ready Clamp is indicated to stop difficult inguinal bleeding on the battlefield, the most common type of junctional bleeding and now the most common cause of preventable battlefield death. The purpose of the present study is to report the data of clamp development to help appliers use it correctly. METHODS Wake Forest University investigators used a cadaver model to test...