Comparing Accuracy of Cervical Pedicle Screw Placement between a Guidance System and Manual Manipulation: A Cadaver Study

نویسندگان

  • Yu Cong
  • Nirong Bao
  • Jianning Zhao
  • Guangping Mao
چکیده

BACKGROUND The aim of this study was to compare the accuracy of cervical pedicle screw placement between a three-dimensional guidance system and manual manipulation. MATERIAL AND METHODS Eighteen adult cadavers were randomized into group A (n=9) and group B (n=9). Ninety pedicle screws were placed into the C3-C7 under the guidance of a three-dimensional locator in group A, and 90 screws were inserted by manual manipulation in group B. The cervical spines were scanned using computed tomography (CT). Parallel and angular offsets of the screws were compared between the two placement methods. RESULTS In group A, 90% of the screws were within the pedicles and 10% breached the pedicle cortex. In group B, 55.6% were within the pedicle and 44.4% breached the pedicle cortex. Locator guidance showed significantly lower parallel and angular offsets in axial CT images (P<0.01), and significantly lower angular offset in sagittal CT images (P<0.01) than manual manipulation. CONCLUSIONS Locator guidance is superior to manual manipulation in accuracy of cervical screw placement. Locator guidance might provide better safety than manual manipulation in placing cervical screws.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of fluoroscopy versus computer-assisted navigation for the placement of anterior cervical pedicle screws.

STUDY DESIGN Randomized laboratory cadaver study. OBJECTIVE The objective of this study was to determine the accuracy of anterior transpedicular screw placement in the cervical spine using conventional fluoroscopy versus computer-assisted navigation. SUMMARY OF BACKGROUND DATA Traditionally, global cervical instability has required anterior and posterior fixation due to the superior biomech...

متن کامل

Cervical Pedicle Screw Placement in Sawbone Models and Unstable Cervical Traumatic Lesions by Using Para-Articular Mini-Laminotomy: A Novice Neurosurgeon’s Experience

Cervical pedicle screw has been shown to provide stronger fixation than other methods. Due to these virtue of biomechanical characteristic, pedicle screw fixation can be used in a wide range of cervical spinal disorders, and excellent clinical results have been reported. However, this procedure is technically demanding because of the great variation in pedicle dimension and angulations between ...

متن کامل

A Novel Patient-Specific Drill Guide Template for Pedicle Screw Insertion into the Subaxial Cervical Spine Utilizing Stereolithographic Modelling: An In Vitro Study

STUDY DESIGN Cadaveric study. PURPOSE The purpose of this study was to assess the accuracy and feasibility of cervical pedicle screw (CPS) insertion into the subaxial cervical spine placed using a patient-specific drill guide template constructed from a stereolithographic model. OVERVIEW OF LITERATURE CPS fixation is an invaluable tool for posterior cervical fixation because of its biomecha...

متن کامل

Assessment of pedicle screw placement accuracy, procedure time, and radiation exposure using a miniature robotic guidance system.

STUDY DESIGN Controlled, cadaveric implantation trial. OBJECTIVE To evaluate the effect of a robotic guidance system on screw placement accuracy, amount of radiation exposure, and length of procedure time during percutaneous pedicle screw implantation. SUMMARY OF BACKGROUND DATA Pedicle screws are associated with low complication rates, and several computer-assisted image guidance systems e...

متن کامل

Placement of pedicle screws in the human cadaveric cervical spine: comparative accuracy of three techniques.

STUDY DESIGN This investigation was conducted in two parts. In the first part, a morphometric analysis of critical cervical pedicle dimensions were measured to create guidelines for cervical pedicle screw fixation based on posterior cervical topography. In the second part of the study, a human cadaver model was used to assess the accuracy and safety of transpedicular screw placement in the suba...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 21  شماره 

صفحات  -

تاریخ انتشار 2015