There is public concern over the long term systemic health effects of metal released from hip replacement prostheses that use large-diameter metal-on-metal bearings. However, to date there has been no systematic study to determine which organs may be at risk, or the magnitude of any effect. We undertook a detailed cross-sectional health screen at a mean of 8 years after surgery in 35 asymptomatic patients who had previously received a metal-on-metal hip resurfacing (MoMHR) versus 35 individually age and sex matched asymptomatic patients who had received a conventional hip replacement. Total body bone mineral density was 5% higher (mean difference 0.05 g/cm², P = 0.02) and bone turnover was 14% lower (TRAP 5b, mean difference -0.56IU/L, P = 0.006; osteocalcin, mean difference -3.08 ng/mL, P = 0.03) in the hip resurfacing versus conventional hip replacement group. Cardiac ejection fraction was 7% lower (mean absolute difference -5%, P = 0.04) and left ventricular end-diastolic diameter was 6% larger (mean difference 2.7 mm, P = 0.007) in the hip resurfacing group versus those patients who received a conventional hip replacement. The urinary fractional excretion of metal was low (cobalt 5%, chromium 1.5%) in patients with MoMHR, but creatinine clearance was normal. Diuretic prescription was associated with a 40% increase in the fractional excretion of chromium (mean difference 0.5%, P = 0.03). There was no evidence of difference in neuropsychological, renal tubular, hepatic or endocrine function between groups (P>0.05). Our findings of differences in bone and cardiac function between patient groups suggest that chronic exposure to low elevated metal concentrations in patients with well-functioning MoMHR prostheses may have systemic effects. Long-term epidemiological studies in patients with well-functioning metal on metal hip prostheses should include musculoskeletal and cardiac endpoints to quantitate the risk of clinical disease.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Distributed analysis of hip implants using six national and regional registries: comparing metal-on-metal with metal-on-highly cross-linked polyethylene bearings in cementless total hip arthroplasty in young patients.

BACKGROUND The regulation of medical devices has attracted controversy recently because of problems related to metal-on-metal hip implants. There is growing evidence that metal-on-metal implants fail early and cause local and systemic complications. However, the failure associated with metal-on-metal head size is not consistently documented and needs to be communicated to patients and surgeons....

متن کامل

Running head: PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES 1 PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES: A LITERTURE REVIEW WITH CASE STUDIES November 13, 2013 PSEUDOTUMORS IN ASSOCIATION WITH HIP PROSTHESES

With the exhibit of pseudotumor discoveries around hip replacements, the search for the cause of such abnormal masses has become a topic of great interest. Possible areas of causation include metal-on-metal hip replacements compared to metal-on-polyethylene hip replacements, patient hypersensitivity to prosthesis debris, or even prosthesis surgical position has been investigated. Although more ...

متن کامل

Auditory and Visual Health after Ten Years of Exposure to Metal-on-Metal Hip Prostheses: A Cross-Sectional Study Follow Up

Case reports of patients with mal-functioning metal-on-metal hip replacement (MoMHR) prostheses suggest an association of elevated circulating metal levels with visual and auditory dysfunction. However, it is unknown if this is a cumulative exposure effect and the impact of prolonged low level exposure, relevant to the majority of patients with a well-functioning prosthesis, has not been studie...

متن کامل

Metallosis of the Resurfaced Hip

Metallosis of the hip is usually defined as aseptic fibrosis, local necrosis, or loosening of the prosthesis secondary to metallic corrosion and release of wear debris.8 It has also been defined as gray discoloration of the tissues of the joint, pain, an effusion, and elevated serum metal levels. Metallosis has been found with stainless steel, titanium, and cobalt chromium alloy femoral prosthe...

متن کامل

Influence of head size on the development of metallic wear and on the characteristics of carbon layers in metal-on-metal hip joints

BACKGROUND AND PURPOSE Particles originating from the articulating surfaces of hip endoprostheses often induce an inflammatory response, which can be related to implant failure. We therefore analyzed the metal content in capsular tissue from 44 McKee-Farrar metal-on-metal hip prostheses (with 3 different head sizes) and we also analyzed the morphological structure of layers located on articulat...