Online conductivity monitoring of dialysis adequacy versus Kt/V derived from urea reduction ratio: A prospective study from a Saudi center

نویسندگان

  • Khalid Al Saran
  • Alaa Sabry
  • Mamdouh Abdulghafour
  • Ahmed Yehia
چکیده

INTRODUCTION Ad equate delivered dose of solute removal (as assessed by urea reduction and calculation of Kt/V) is an important determinant of clinical outcome in chronic hemodialysis (HD) patients. This requires both prescription of an adequate dose of HD and regular assessment that the delivered treatments are also adequate. Online conductivity monitoring using sodium flux as a surrogate for urea allows the repeated noninvasive measurement of Kt/V on each HD treatment. METHODS We prospectively studied 17 (9 males, 8 females) established chronic HD patients over an eight-week period (408 treatments). A pre- and post-dialyzer measurement of the conductivity is performed by two mutually independent temperature-compensated conductivity cells equipped with Fresenius 4008 S(®) dialysis machines. Urea reduction was measured (once a week) by a single pool calculation using immediate post-treatment sampling. No changes were made to any of the dialysis prescriptions over the study period. Values of calculated Kt/V and simultaneously obtained online Kt/V were compared. RESULTS There was a statistically significant difference between calculated Kt/V and online Kt/V over the study period. The mean calculated Kt/V was 1.37 ± 0.09, and mean online Kt/V 1.02 ± 0.15 (P = 0.000), calculated Kt/V ≥ 1.2 was achieved in all our patients while online Kt/V ≥ 1.2 was achieved in only 17.64 %. Yet there was moderate correlation between calculated Kt/V and online Kt/V (r(2) = 0.48). CONCLUSIONS Online conductivity monitoring results underestimates dialysis efficiency compared to calculated Kt/V readings. This difference has to be considered when applying Kt/V to clinical practice.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

بررسی کفایت دیالیز در بیماران همودیالیزی مرکز آموزشی درمانی رازی قائم شهر در سال 1391

Background and purpose: Hemodialysis is the most common treatment that is used in end-stage renal disease. Since there is not a clear picture of how the adequacy of hemodialysis treatment at this center, this study aimed to determine the adequacy of dialysis in hemodialysis patients at Qaemshar Razi Hospital in 2012. Material and methods: This was a cross-sectional study. The samples were sele...

متن کامل

تاثیر بولوس متناوب محلول نرمال سالین در حین همودیالیز بر کفایت دیالیز

  Background and purpose: Using practical and non-invasive methods to improve the adequacy of dialysis (i.e. Kt/V) can increase the quality of life and decrease the mortality and morbidity rate in patients under hemodialysis. This study was conducted to investigate the effect of intermittent normal saline bolus on adequacy of dialysis. Materials and methods: In this randomized controlled...

متن کامل

Assessment of hemodialysis adequacy in patients undergoing maintenance maneuver by laboratory tests

Background and Objectives: Hemodialysis (HD) is one of the therapeutic modalities patients. Since inadequate dialysis increases mortality and morbidity of patients, therefore, assessment of dialysis adequacy is clinically important. For this reason, hemodialysis adequacy was determined in patients underg...

متن کامل

بررسی کفایت دیالیز در بیماران همودیالیزی بیمارستان شهید آیت الله اشرفی اصفهانی شهر تهران در سال 1395

Background &Objective: The inadequacy of dialysis is one of the determinants of inability in hemodialysis patients. Due to the lack of clarity of the adequacy of dialysis in hemodialysis patients in Shahid Ayatollah Ashrafi Esfahani Hospital in Tehran, and since there has not been a similar study in this area, this study was the first study to evaluate the adequacy of dialysis in patients under...

متن کامل

بررسی کفایت دیالیز در بیماران همودیالیزی مراجعه کننده به مرکز همودیالیز شهر بوشهر

Introduction & Objective: When hemodialysis treatment was first used to replace renal function, no one knew how mach dialysis therapy was needed to keep the patients healthy. Dialysis efficacy is one of the predominant factors survivals in hemodialysis patients. There are multiple methods for assessing the adequacy of dialysis. The index KT/V, most commonly used marker for dialysis adequacy. Th...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2  شماره 

صفحات  -

تاریخ انتشار 2009