Determination of Aluminum in Dialysis Concentrates by Atomic Absorption Spectrometry after Coprecipitation with Lanthanum Phosphate
نویسندگان
چکیده
This method was developed for the determination of trace amounts of aluminum(III) in dialysis concentrates using atomic absorption spectrometry after coprecipitation with lanthanum phosphate. The analytical parameters that influenced the quantitative coprecipitation of analyte including amount of lanthanum, amount of phosfate, pH and duration time were optimized. The % recoveries of the analyte ion were in the range of 95-105 % with limit of detection (3s) of 0.5 µg l-1. Preconcentration factor was found as 1000 and Relative Standard Deviation (RSD) % value obtained from model solutions was 2.5% for 0.02 mg L-1. The accuracy of the method was evaluated with standard reference material (CWW-TMD Waste Water). The method was also applied to most concentrated acidic and basic dialysis concentrates with satisfactory results.
منابع مشابه
Determination of Aluminum in Dialysis Concentrates by Atomic Absorption Spectrometry after Coprecipitation with Lanthanum Phosphate
This method was developed for the determination of trace amounts of aluminum(III) in dialysis concentrates using atomic absorption spectrometry after coprecipitation with lanthanum phosfate. The analytical parameters that influenced the quantitave coprecipitation of analyte including amount of lanthanum, amount of phosfate, pH and duration time were optimized. The % recoveries of the analyte io...
متن کاملDetermination of Aluminum in Dialysis Concentrates by Atomic Absorption Spectrometry after Coprecipitation with Lanthanum Phosphate
This method was developed for the determination of trace amounts of aluminum(III) in dialysis concentrates using atomic absorption spectrometry after coprecipitation with lanthanum phosfate. The analytical parameters that influenced the quantitave coprecipitation of analyte including amount of lanthanum, amount of phosfate, pH and duration time were optimized. The % recoveries of the analyte io...
متن کاملSeparation/preconcentration of trace heavy metals in urine, sediment and dialysis concentrates by coprecipitation with samarium hydroxide for atomic absorption spectrometry.
Multi-element determination of trace elements in urine and dialysis solutions by atomic absorption spectrometry has been investigated. Coprecipitation with samarium hydroxide was used for preconcentration of trace elements and elimination of matrix elements. To 10 ml of each sample was added 500 mul of 2 mgml(-1) samarium solutions; the pH was then adjusted to 12.2 in order to collect trace hea...
متن کاملDetermination of cadmium in river water by electrothermal atomic absorption spectrometry after internal standardization-assisted rapid coprecipitation with lanthanum phosphate.
Cadmium ranging from 1 - 8 ng could be coprecipitated quantitatively with lanthanum phosphate at pH 5 - 6 from up to 200 mL of river water samples spiked with 5 microg of indium as an internal standard. Cadmium and indium coprecipitated were measured by using electrothermal atomic absorption spectrometry. The cadmium content in the original sample solution could be determined by internal standa...
متن کاملApplication of internal standardization to rapid coprecipitation technique using lanthanum phosphate for flame atomic absorption spectrometric determination of iron and lead.
By applying an internal standardization, we could use a rapid coprecipitation technique using lanthanum phosphate as a coprecipitant for preconcentration of iron(III) and lead in their flame atomic absorption spectrometric determination. Indium as an internal standard was added to the initial sample solution together with lanthanum and phosphoric acid; the coprecipitation of iron(III) and lead ...