RADIATION PROTECTION OF PATIENT DURING STEREOTACTIC RADIOSURGERY. Part I: Single isocenter phantom measurements

نویسنده

  • MUSTAFA MAJALI
چکیده

Radiation dose to organs outside the radiotherapy treatment field can be significant and clinical interest. Therefore, in order to estimate peripheral doses (PD) during the stereotactic radiosurgery treatment with Leksell Gamma Knife for single isocenter treatment technique in particular as a part of the method can be described the all plan treatment in general. Knowledge of PD can be used effectively during the optimization of treatment planning in order to deliver a very accurate dose of radiation to well-defined target with minimal damage to surrounding healthy tissue and other organs. Moreover, PD can be limiting factor during the treatment of some non-malignant lesion or pregnant woman. Alderson-Rando phantom was used for PD measurements. The Dose of 40 Gy was delivered in a single fraction radiosurgery to midline hypothetical target volume situated close to the center of calva. Treatment planning for single target covered by 50% isodose of each collimator was performed with Leksell Gamma Plan (ver. 5,32) treatment planning system. The treatment plans for a single, 5, 10, 15, 20 and 25 shots, keeping the same position and the same doses were produced. Automatic positioning system (APS) was used for positioning of phantom during the treatment. The doses to different organs were measured during treatment by thermoluminescence dosimeter (TL). A simple exponential function can describe dependence of measured PD on distance from target for all collimator size were utilizing. Close to the target a scattering radiation within the patient was dominant, at distances beyond 40 cm from the target this component becomes less predominated and less dependence of the collimator size. While that the PD distances beyond 40 cm was mainly due to scattered radiation in the treatment room and leakage radiation. Contributions (DT) to the PD due the transportation of patient during the positioning were linearly increasing with the number of shots. DT does not practically contribute to the prescribed dose in target and to the laterally scattered radiation within patient. PD depends on collimator size and this dependence is more significant close to the target where it is proportional with the C3 and less significant far from the target where it is proportional with C only but more dependence on the number of shots. The number of shots as well as collimator size has influence on the magnitude of PD. Reduction of number of shots can decrease the PD for sites far from the target, on other hand the influence of collimator size is more significant for organs close to the target. Weighting the effect of the collimator size and number of shots can be used effectively during the optimization procedure to choose the most suitable treatment plan. 1 Faculty of Nuclear Sciences and Physical Engineering, Brehova 7, 115 19, Prague 1, Prague, Czech Republic.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Development of a prototype stereotactic collimation assembly for Neptun 10 PC linac

Background: A set of hardware stereotactic radiosurgery is designed and constructed. The scope of this article is to describe how we have designed and developed stereotactic collimation assembly for a 9 MV Neptun 10 PC linac. Materials and methods: One collimator holder and six collimators were made, with their radiation field diameter between 12.5 mm to 25 mm at isocenter level. Beam data, suc...

متن کامل

Measurement of stereotactic beam data of 9 MV X-rays using diode detector

Background: Radiosurgery is a focal brain irradiation technique that delivers, usually in a single session, high dose of ionizing radiation. The presence of lateral electronic disequilibrium and steep dose gradients in small fields demands special attention to the selection of a suitable detector with respect to its size, composition and response. Materials and Methods : Small circular fields w...

متن کامل

Pretreatment setup verification by cone beam CT in stereotactic radiosurgery: phantom study

Kilovoltage cone beam computed tomography (CBCT) imaging may be useful in verifying patient position in stereotactic radiosurgery (SRS). To evaluate its efficacy, we investigated isocenter differences in the radiation beam and CBCT with respect to the achievable setup of a conventional frame-based SRS system. A verification phantom constructed from two plastic boards and Gafchromic-EBT film (4 ...

متن کامل

SU-F-T-547: Off-Isocenter Winston-Lutz Test for Stereotactic Radiosurgery/stereotactic Body Radiotherapy.

PURPOSE To perform a quantitative study to verify that the mechanical field center coincides with the radiation field center when both are off from the isocenter during the single-isocenter technique in linear accelerator-based SRS/SBRT procedure to treat multiple lesions. METHODS We developed an innovative method to measure this accuracy, called the off-isocenter Winston-Lutz test, and here ...

متن کامل

Quality Assurance and Commissioning of an Infrared Marker-Based Patient Positioning System for Frameless Extracranial Stereotactic Radiotherapy

Rapid advancements in imaging technology have led to remarkable improvements in identification and localization of tumors, ushering the era of high-precision techniques in contemporary radiotherapy practice. However, uncertainties in patient set-up and organ motion during a course of fractionated radiotherapy can compromise precision of radiation therapy. Excellent accuracy has been achieved wi...

متن کامل