A Recommendation for Revised Dose Calibrator Measurement Procedures for 89Zr and 124I
نویسندگان
چکیده
Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89)Zr the PET images should, in principle, be as quantitatively accurate as similarly reconstructed 18F measurements. We found, however, that images of a 20 cm well calibration phantom containing (89)Zr underestimated the activity by approximately 10% relative to a dose calibrator measurement (Capintec CRC-15R) using a published calibration setting number of 465. PET images of (124)I, in contrast, are complicated by the contribution of decays in cascade that add spurious coincident events to the PET data. When these cascade coincidences are properly accounted for, quantitatively accurate images should be possible. We found, however, that even with this correction we still encountered what appeared to be a large variability in the accuracy of the PET images when compared to dose calibrator measurements made using the calibration setting number, 570, recommended by Capintec. We derive new calibration setting numbers for (89)Zr and (124)I based on their 511 keV photon peaks as measured on an HPGe detector. The peaks were calibrated relative to an 18F standard, the activity level of which was precisely measured in a dose calibrator under well-defined measurement conditions. When measuring (89)Zr on a Capintec CRC-15R we propose the use of calibration setting number 517. And for (124)I, we recommend the use of a copper filter surrounding the sample and the use of calibration setting number 494. The new dose calibrator measurement procedures we propose will result in more consistent and accurate radioactivity measurements of (89)Zr and (124)I. These and other positron emitting radionuclides can be accurately calibrated relative to 18F based on measurements of their 511 keV peaks and knowledge of their relative positron abundances.
منابع مشابه
aHER2 Fab with Optimized Plasma Half-Life for High-Contrast Tumor Imaging In Vivo
Immuno-PET imaging of the tumor antigen HER2/neu allows for the noninvasive detection and monitoring of oncogene expression; such detection and monitoring are of prognostic value in patients with breast cancer. Compared with the full-size antibody trastuzumab, smaller protein tracers with more rapid blood clearance permit higher imaging contrast at earlier time points. Antigen-binding fragments...
متن کاملBiokinetic modelling of 89-Zr-labelled monoclonal antibodies for dosimetry assessment in humans
Background: Monoclonal antibodies have confirmed their merit as biotherapeutics across a wide spectrum of diseases, including cancer, heart disease, infection, and immune disorders. Materials and Methods: The dynamics of 89Zr-labelled monoclonal antibodies (MAb) after injection into the human body are modelled. This modified biokinetic model can be used for dose assessment not only for 89Zr-lab...
متن کاملMulticentre harmonization of 89Zr PET/CT performance
This study investigated the feasibility of quantitative accuracy and harmonized image quality in Zr-PET/CT multicentre studies. Methods: Five PET/CT scanners from 3 vendors were included. Zr activity was measured in a central dose calibrator before delivery.Local activity assays were based on volume as well as on the local dose calibrator. Accuracy and image noise were determined from a cross c...
متن کاملEvaluation of Maximum Patient Skin Dose Arising from Interventional Cardiology Using Thermoluminescence Dosimeter in Mashhad, Iran
Introduction The increasing practice of interventional fluoroscopy in diagnosis and treatment of cardiovascular disease has risen attention to improve radiation protection of patients and cardiologists in these relatively high dose techniques. Therefore, nowadays there is an emphasis on the measurement of radiation dose received by patients and cardiologists arising from the relevant procedures...
متن کاملMulticenter harmonization of 89Zr PET/CT performance.
UNLABELLED This study investigated the feasibility of quantitative accuracy and harmonized image quality in (89)Zr-PET/CT multicenter studies. METHODS Five PET/CT scanners from 3 vendors were included. (89)Zr activity was measured in a central dose calibrator before delivery. Local activity assays were based on volume as well as on the local dose calibrator. Accuracy and image noise were dete...
متن کامل