Correction of motion‐induced misalignment in co‐registered PET/CT and MRI (T1/T2/FLAIR) head images for stereotactic radiosurgery†
نویسندگان
چکیده
The purpose was to evaluate and correct the co-registration of diagnostic PET/CT and MRI/MRI images for stereotactic radiosurgery (SRS) using 3D volumetric image registration (3DVIR). The 3DVIR utilizes the homogeneity of color distribution over a volumetric anatomical landmark as the registration criterion with submillimeter accuracy. Fifty-three PET/CT and MRI (T1, T2 and FLAIR) image sets of patients with brain lesions were acquired sequentially from a hybrid PET/CT or an MRI scanner with common diagnostic head holding devices. Twenty-five sets of head 18F-FDG-PET/CT images were scanned over a 10-minute interval and 14 whole-body sets were scanned over a 30-minute interval. Fourteen sets of MRI images were acquired, and each 3-modal image set (T1, T2 and FLAIR) was scanned in sequence at time 0, ~5 and ~20 minutes. The misalignments in these "co-registered" images were evaluated and corrected using the 3DVIR. Using the head immobilization devices commonly found in diagnostic PET/CT and MRI/MRI imaging, 80%-100% of these "co-registered" images were identified as misaligned. For PET/CT, the magnitude of misalignment was 0.4° ± 0.5° and 0.7 ± 0.4 mm for 10-minute scans, and 0.8° ± 1.2° and 2.7 ± 1.7 mm for 30-minute scans. For MRI/MRI, the magnitude was 0.2° ± 0.4° and 0.3 ± 0.2 mm for 5-minute scan intervals, and 1.1° ± 0.7° and 1.2 ± 1.4 mm for 20-minute intervals. Small, but significant, misalignment is present in the co-registered diagnostic PET/CT and MRI/MRI images and can be corrected in SRS treatment planning using the volumetric image registration for improved target localization within the clinical error tolerance.
منابع مشابه
Design, Construction and Evaluation of an Anthropomorphic Head Phantom for Assessment of Image Distortion in Stereotactic Radiosurgery Planning Systems
Introduction: In recent years, the use of magnetic resonance (MR) images in radiation treatment planning has drawn considerable attention. However, although the extent of a tumor can be determined in great detail on MR images, the geometric accuracy of these images is limited by distortions stemming from the inhomogeneity of the static background magnetic field, the nonlineari...
متن کاملReliability of Stereotactic Coordinates of 1.5-Tesla and 3-Tesla MRI in Radiosurgery and Functional Neurosurgery
OBJECTIVE The aims of this study are to identify interpersonal differences in defining coordinates and to figure out the degree of distortion of the MRI and compare the accuracy between CT, 1.5-tesla (T) and 3.0T MRI. METHODS We compared coordinates in the CT images defined by 2 neurosurgeons. We also calculated the errors of 1.5T MRI and those of 3.0T. We compared the errors of the 1.5T with...
متن کاملValidation of accuracy in image co-registration with computed tomography and magnetic resonance imaging in Gamma Knife radiosurgery
The latest version of Leksell GammaPlan (LGP) is equipped with Digital Imaging and Communication in Medicine (DICOM) image-processing functions including image co-registration. Diagnostic magnetic resonance imaging (MRI) taken prior to Gamma Knife treatment is available for virtual treatment pre-planning. On the treatment day, actual dose planning is completed on stereotactic MRI or computed to...
متن کاملImpact of PET - CT motion correction in minimising the gross tumour volume in non-small cell lung cancer
AbstractObjective: To investigate the impact of respiratory motion on localization, and quantification lung lesions for the Gross Tumour Volume utilizing an in-house developed Auto3Dreg programme and dynamic NURBS-based cardiac-torso digitised phantom (NCAT). Methods: Respiratory motion may result in more than 30% underestimation of the SUV values of lung, liver and kidney tumour lesions. The m...
متن کاملUse of stereotactic PET images in dosimetry planning of radiosurgery for brain tumors: clinical experience and proposed classification.
UNLABELLED We developed a technique that allows the routine integration of PET in stereotactic neurosurgery, including radiosurgery. We report our clinical experience with the combined use of metabolic (i.e., PET) and anatomic (i.e., MRI and CT) images for the radiosurgical treatment of brain tumors. We propose a classification describing the relative role of the information provided by PET in ...
متن کامل