Rectal protection in prostate stereotactic radiotherapy: a retrospective exploratory analysis of two rectal displacement devices
نویسندگان
چکیده
INTRODUCTION High rectal doses are associated with increased toxicity. A rectal displacement device (RDD) reduces rectal dose in prostate stereotactic body radiation therapy (SBRT). This study investigates any dosimetric difference between two methods of rectal displacement (Rectafix and SpaceOAR) for prostate SBRT. METHODS Rectal dosimetry of 45 men who received SBRT within the PROMETHEUS trial was retrospectively examined, across two radiation therapy centres using the two RDD's. Men received a total dose (TD) of 19 or 20 Gy in two fractions followed by 46 Gy in 23 fractions. Centre 1 contributed 16 Rectafix and 10 SpaceOAR patients. Centre 2 contributed 19 Rectafix patients. Rectal dose volume histogram (DVH) data were recorded as a TD percentage at the following volume intervals; V1%, V2%, V5%, V10% and then 10% increments to V80%. As only one centre employed both RDD's, three sequential rectal dosimetry comparisons were performed; (1) centre 1 Rectafix versus centre 1 SpaceOAR; (2) centre 1 Rectafix versus centre 2 Rectafix and (3) centre 1+ centre 2 Rectafix versus centre 1 SpaceOAR. RESULTS In comparison (1) Rectafix demonstrated lower mean doses at 9 out of 11 measured intervals (P = 0.0012). Comparison (2) demonstrated a moderate difference with centre 2 plans producing slightly lower rectal doses (P = 0.013). Comparison (3) further demonstrated that Rectafix returned lower mean doses than SpaceOAR (P < 0.001). Although all dose levels were in favour of Rectafix, in absolute terms differences were small (2.6-9.0%). CONCLUSIONS In well-selected prostate SBRT patients, Rectafix and SpaceOAR RDD's provide approximately equivalent rectal sparing.
منابع مشابه
The effect of Rectal Retractor, a rectal protector on rectal dose in radical prostate radiotherapy
Introduction: The close proximity of the rectum tissue to the prostate gland results in rectal toxicity during prostate radiotherapy. Rectal Retractor (RR) is a rectal displacement technique that inserts into the rectum, as a result, the lateral and dorsal part of the rectal wall push away from high dose areas. The aim of this study was to investigate the dosimetric consequence...
متن کاملReal‐time intrafraction prostate motion during linac based stereotactic radiotherapy with rectal displacement
BACKGROUND Kilovoltage Intrafraction Monitoring (KIM) is a method which determines the three-dimensional position of the prostate from two-dimensional kilovoltage (kV) projections taken during linac based radiotherapy treatment with real-time feedback. Rectal displacement devices (RDDs) allow for improved rectal dosimetry during prostate cancer treatment. This study used KIM to perform a prelim...
متن کاملReal-time in vivo rectal wall dosimetry using MOSkin detectors during linac based stereotactic radiotherapy with rectal displacement
BACKGROUND MOSFET dosimetry is a method that has been used to measure in-vivo doses during brachytherapy treatments and during linac based radiotherapy treatment. Rectal displacement devices (RDDs) allow for safe dose escalation for prostate cancer treatment. This study used dual MOSkin detectors to assess real-time in vivo rectal wall dose in patients with an RDD in place during a high dose pr...
متن کاملProstate IMRT: Two-dimensional model of rectal NTCP employing the variability of rectal motion and rectum wall thickness
Background: In order to improve the evaluation of possible rectal toxicity based on the rectal normal tissue complication probability (NTCP), we consider the fractional dependence of the NTCP on the wall thickness (tW) and rectal displacement (RM). Materials and Methods: The two-dimensional NTCP model (NTCP2D) was developed using radiotherapy plans of ten randomly selected patients with prostat...
متن کاملRectal endoscopy findings following stereotactic body radiation therapy for clinically localized prostate cancer
BACKGROUND Treating prostate cancer with SBRT could potentially minimize radiation proctitis by reducing high-dose rectal irradiation. In addition, it offers the potential radiobiologic benefits of hypofractionation. This study reports the endoscopic changes and the associated clinical rectal toxicity in these patients. METHODS We reviewed the records of patients treated from 2008-2011 for lo...
متن کامل