Reprocessing endoscopes: United States perspective.

نویسندگان

  • W A Rutala
  • D J Weber
چکیده

Endoscopes are used frequently for the diagnosis and therapy of medical disorders. For example, greater than 10000000 gastrointestinal endoscopic procedures are performed each year in the United States. Failure to employ appropriate cleaning and disinfection/sterilization of endoscopes has been responsible for multiple nosocomial outbreaks and serious, sometimes life-threatening, infections. Flexible endoscopes, by virtue of the site of use, have a high bioburden of microorganisms after use. The bioburden found on flexible gastrointestinal endoscopes following use has ranged from 10(5) to 10(10)CFU/ml, with the highest levels being found in the suction channels. Cleaning dramatically reduces the bioburden on endoscopes. Several investigators have shown a mean log(10) reduction factor of 4 (99.99%) in the microbial contaminants with cleaning alone. Cleaning should be done promptly following each use of an endoscope to prevent drying of secretions, allow removal of organic material, and decrease the number of microbial pathogens. Because the endoscope comes into intimate contact with mucous membranes, high-level disinfection is the reprocessing standard after each patient use. High-level disinfection refers to the use of a disinfectant (e.g., FDA-cleared chemical sterilant or high-level disinfectant) that inactivates all microorganisms (i.e., bacteria, viruses, fungi, mycobacteria) but not high levels of bacterial spores. The disinfection process requires immersion of the endoscope in the high-level disinfectant and ensuring all channels are perfused for the approved contact time (e.g., for ortho-phthaladehyde this is 12 min in the US). Following disinfection, the endoscope and channels are rinsed with sterile water, filtered water, or tapwater. The channels are then flushed with alcohol and dried using forced air. The endoscope should be stored in a manner that prevents recontamination. A protocol that describes the meticulous manual cleaning process, the appropriate training and evaluation of the reprocessing personnel, and a quality assurance program for endoscopes should be adopted and enforced by each unit performing endoscopic reprocessing.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Computer-aided recording of automatic endoscope washing and disinfection processes as an integral part of medical documentation for quality assurance purposes

BACKGROUND The reprocessing of medical endoscopes is carried out using automatic cleaning and disinfection machines. The documentation and archiving of records of properly conducted reprocessing procedures is the last and increasingly important part of the reprocessing cycle for flexible endoscopes. METHODS This report describes a new computer program designed to monitor and document the auto...

متن کامل

Multi-society guideline for reprocessing flexible gastrointestinal endoscopes.

Flexible gastrointestinal endoscopy is a valuable diagnostic and therapeutic tool for the care of patients with gastrointestinal and pancreaticobiliary disorders. Compliance with accepted guidelines for the reprocessing of gastrointestinal endoscopes between patients is critical to the safety and success of their use. When these guidelines are followed, pathogen transmission can be effectively ...

متن کامل

Residual bioburden in reprocessed side-view endoscopes used for endoscopic retrograde cholangiopancreatography (ERCP)

BACKGROUND AND STUDY AIM Worldwide some endoscopy units routinely continue to use manual reprocessing techniques for disinfection of side-view endoscopes. The aim of this study was to evaluate the outcome quality of manual reprocessing techniques for removal and inactivation of the bioburden from side-view endoscopes used for endoscopic retrograde cholangiopancreatography (ERCP) in a tertiary r...

متن کامل

Recent advances in epidemiology and prevention of gastrointestinal endoscopy related infections.

PURPOSE OF REVIEW This article reviews recent publications relevant to endoscope reprocessing and the potential for transmission of infection during gastrointestinal endoscopy. RECENT FINDINGS There have been a number of established reprocessing failures of gastrointestinal endoscopes at various healthcare facilities across the US resulting in patient notifications. These episodes have been a...

متن کامل

WHO EMRO | Reprocessing practices for gastrointestinal endoscopes: a multicentre study in Egyptian university hospitals PDF version Short communication

The aim of this study was to assess the practices of health care workers during gasterointestinal endoscope reprocessing, evaluate their knowledge about reprocessing, and verify their compliance with laboratory and microbiological tests in endoscopy units at Zagazig University and Fayoum University hospitals. All nursing staff on duty from 10 endoscopy units, with 16 flexible endoscopes, were i...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of hospital infection

دوره 56 Suppl 2  شماره 

صفحات  -