Nebulized albuterol delivery in a model of spontaneously breathing children with tracheostomy.

نویسنده

  • Ariel Berlinski
چکیده

BACKGROUND Nebulized therapy is commonly used in spontaneously breathing tracheostomized patients, despite a lack of recommended devices and techniques. I compared albuterol dose delivered to a model of spontaneously breathing children with tracheostomy, using different nebulizers, tracheostomy tube sizes, inhalation techniques, and breathing patterns. METHODS A tracheostomy model was connected in series to a breathing simulator, with a filter interposed. I simulated the breathing patterns of a 16-month-old child and 12-year-old child, and tested tracheostomy tubes with internal diameters of 3.5 mm and 5.5 mm. Albuterol nebulizer solution (2.5 mg/3 mL) was used. A breath-enhanced nebulizer (Pari LC Plus), a breath-actuated nebulizer (AeroEclipse), and a nebulizer that continuously delivers aerosol (Up-Draft II Opti-Neb) were operated for 5 min at 6 L/min with wall air. The Up-Draft II was tested with T-piece and mask interfaces, with an extension tube, and with and without assisted breathing (every breath and every other breath). The amount of albuterol delivered was analyzed via spectrophotometry. Particle size distribution was measured with a cascade impactor. RESULTS The Pari LC Plus was more efficient than the Up-Draft II or AeroEclipse. Assisted breathing with the Up-Draft II with extension increased albuterol delivery with every other breath (second best device/configuration), being superior to every breath technique. Adding an extension tube increased delivered albuterol. T-piece was more efficient than mask. Breathing patterns with larger tidal volume increased albuterol delivery. Tracheostomy size had less impact on drug delivery. Mass median aerodynamic diameter decreased by 48-74% when passing through the tracheostomy tubes, and 0.8% of the nominal dose was deposited in the tracheostomy tube. CONCLUSIONS Albuterol delivery in a model of spontaneously breathing children with tracheostomy is influenced by type of device and configuration, use of assisted breathing, breathing pattern, and tracheostomy tube size. Mass median aerodynamic diameter significantly decreases during passage through a tracheostomy tube.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Albuterol Delivery via Facial and Tracheostomy Route in a Model of a Spontaneously Breathing Child.

BACKGROUND Some pediatric patients receiving therapeutic aerosols undergo tracheostomy, and others who are tracheostomized continue requiring inhaled therapies upon decannulation. It is unknown whether a dose adjustment is required. Different devices are available for facial and tracheostomy delivery, and in some instances, the assisted technique is used. We hypothesized that the change from fa...

متن کامل

Effect of mask dead space and occlusion of mask holes on delivery of nebulized albuterol.

BACKGROUND Infants and children with respiratory conditions are often prescribed bronchodilators. Face masks are used to facilitate the administration of nebulized therapy in patients unable to use a mouthpiece. Masks incorporate holes into their design, and their occlusion during aerosol delivery has been a common practice. Masks are available in different sizes and different dead volumes. The...

متن کامل

Effect of face mask design on inhaled mass of nebulized albuterol, using a pediatric breathing model.

BACKGROUND Aerosol face mask design and the distance at which the face mask is held from the face affect the delivery of nebulized medication to pediatric patients. OBJECTIVE To measure the inhaled mass of nebulized albuterol with 3 types of pediatric face mask, at 3 different distances from the face, with a model of a spontaneously breathing infant. METHODS We compared a standard pediatric...

متن کامل

An in vitro evaluation of aerosol delivery through tracheostomy and endotracheal tubes using different interfaces.

BACKGROUND Previous research reporting factors influencing aerosol delivery in intubated patients has been largely focused on the endotracheal tube (ETT) during mechanical ventilation, with little comparative analysis of effect of types of artificial airways and their interfaces on aerosol delivery during spontaneous breathing. The purpose of this study was to compare aerosol delivery via trac...

متن کامل

Quantifying Aerosol Delivery in Simulated Spontaneously Breathing Patients With Tracheostomy Using Different Humidification Systems With or Without Exhaled Humidity.

BACKGROUND Aerosol and humidification therapy are used in long-term airway management of critically ill patients with a tracheostomy. The purpose of this study was to determine delivery efficiency of jet and mesh nebulizers combined with different humidification systems in a model of a spontaneously breathing tracheotomized adult with or without exhaled heated humidity. METHODS An in vitro mo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 58 12  شماره 

صفحات  -

تاریخ انتشار 2013