Assessment of the Quantitative Ability of AdvanSure TB/NTM Real-Time PCR in Respiratory Specimens by Comparison with Phenotypic Methods
نویسندگان
چکیده
Accurate quantification of mycobacterial load is important to evaluate disease severity and to monitor the course of treatment in tuberculosis (TB). We evaluated the quantitative capability of the AdvanSure TB/NTM real-time PCR kit (LG Life Science, Korea) to determine the cycle threshold (Ct) for mycobacterial burden. We retrospectively analyzed data from 108 patients whose respiratory specimens (sputums and bronchoalveolar lavage fluids) were positive for Mycobacterium tuberculosis complex (85 culture-positive and 23 culture-negative specimens). We compared Ct values with grades of acid-fast bacilli (AFB) staining, semi-quantitative colony count on solid medium, and time to positivity (TTP) in liquid and solid media. We also investigated the cutoff Ct value for predicting stain-positive status. Ct value showed significant reverse correlation with AFB staining grade (r s=-0.635, P<0.01). Ct value significantly decreased as the semi-quantitative counts on the solid medium increased (P<0.001), and the mean Ct value of each of the groups 1+, 2+, 3+, and 4+ were 29.0, 30.0, 27.1, and 25.5, respectively. A weak correlation between Ct value and TTP in liquid and solid media was observed (r s=0.468 and 0.365, respectively). A cutoff Ct value of <33.2 best predicted stain positivity, with a sensitivity of 95.0% and a specificity of 32.0%. Our findings suggest the potential use of AdvanSure TB/NTM real-time PCR kit for quantitatively determining bacterial burden, albeit with some enhancements.
منابع مشابه
[Evaluation of the performances of AdvanSure TB/NTM real time PCR kit for detection of mycobacteria in respiratory specimens].
BACKGROUND PCR is a widely used method for rapid and accurate diagnosis of mycobacteriosis. The sensitivity and specificity of a real time PCR kit newly developed in Korea were evaluated for detecting mycobacteria in respiratory specimens. METHODS One hundred twenty nine Mycobacterium tuberculosis (TB) culture positive respiratory specimens (82 AFB stain positive and 47 stain negative specime...
متن کاملComparison of AdvanSure TB/NTM PCR and COBAS TaqMan MTB PCR for Detection of Mycobacterium tuberculosis Complex in Routine Clinical Practice
The AdvanSure tuberculosis/non-tuberculous mycobacterium (TB/NTM) PCR (LG Life Science, Korea) and COBAS TaqMan Mycobacterium tuberculosis (MTB) PCR (Roche Diagnostics, USA) are commonly used in clinical microbiology laboratories. We aimed to evaluate these two commercial real-time PCR assays for detection of MTB in a large set of clinical samples over a two-year period. AdvanSure TB/NTM PCR an...
متن کاملComparison of Ogawa Media, BACTEC MGIT 960 System and TB/NTM Real-Time PCR for Detecting Mycobacterium Species
Background: Mycobacterial infection is a problem throughout the world along with the increase of immunocompromised patients. For this reason, there have been many methods for faster and more accurate diagnosis. In this study, we evaluated several laboratory methods for mycobacterial infection. Methods: From January to December 2009, 635 specimens were cultured with mycobacteria growth indicator...
متن کاملPerformance Evaluation of Anyplex plus MTB/NTM and AdvanSure TB/NTM for the Detection of Mycobacterium tuberculosis and Nontuberculous Mycobacteria
Background: Polymerase chain reaction (PCR) methods from direct specimen are widely used for the rapid and accurate detection of mycobacteria infection. In this study, we evaluated two domestically developed detection kits for Mycobacterium tuberculosis (MTB) and nontuberculous mycobacteria (NTM) using real-time PCR. Methods: A total of 348 samples from patients with suspected tuberculosis were...
متن کاملComparison of Diagnostic Performance of Three Real-Time PCR Kits for Detecting Mycobacterium Species
PURPOSE PCR is widely used for rapidly and accurately detecting Mycobacterium species. The purpose of this study was to assess the diagnostic performance of three real-time PCR kits and evaluate the concordance with two older PCR methods. MATERIALS AND METHODS Using 128 samples, the five PCR methods were assessed, including an in-house PCR protocol, the COBAS Amplicor MTB, the COBAS TaqMan MT...