Adapting a generic BEAMnrc model of the BrainLAB m3 micro-multileaf collimator to simulate a local collimation device.
نویسندگان
چکیده
This work is focussed on developing a commissioning procedure so that a Monte Carlo model, which uses BEAMnrc's standard VARMLC component module, can be adapted to match a specific BrainLAB m3 micro-multileaf collimator (microMLC). A set of measurements are recommended, for use as a reference against which the model can be tested and optimized. These include radiochromic film measurements of dose from small and offset fields, as well as measurements of microMLC transmission and interleaf leakage. Simulations and measurements to obtain microMLC scatter factors are shown to be insensitive to relevant model parameters and are therefore not recommended, unless the output of the linear accelerator model is in doubt. Ultimately, this note provides detailed instructions for those intending to optimize a VARMLC model to match the dose delivered by their local BrainLAB m3 microMLC device.
منابع مشابه
The influence of field size on stopping‐power ratios in‐ and out‐of‐field: quantitative data for the BrainLAB m3 micro‐multileaf collimator
The objective of this work is to quantify the systematic errors introduced by the common assumption of invariant secondary electron spectra with changing field sizes, as relevant to stereotactic radiotherapy and other treatment modes incorporating small beam segments delivered with a linac-based stereotactic unit. The EGSnrc/BEAMnrc Monte Carlo radiation transport code was used to construct a d...
متن کاملRadiation transmission, leakage and beam penumbra measurements of a micro‐multileaf collimator using GafChromic EBT film
Micro-multileaf collimator systems coupled to linear accelerators for radioneurosurgery treatments require a rigorous dosimetric characterization in order to be used in 3D conformal and intensity modulated stereotactic radiosurgery and radiotherapy applications. This characterization involves high precision measurements of leaf transmission, leakage and beam penumbra through the collimation sys...
متن کاملStatic and dynamic beam intensity modulation in radiotherapy using a multileaf collimator
Cover: Generation of an intensity modulated field by using dynamic multileaf collimation.
متن کاملEvaluation of MLC Thickness and Composite Effects on Collimation Parameters using EGSnrc and IAEA Phase Space Data
Background: Recently, multileaf collimators (MLC) have become an important part of any LINAC collimation system because they reduce the treatment planning time and improve the conformity. Important factors that affect MLCs collimation performance are leaves material composition and their thickness.Objective: In this study, we investigate main dosimetric parameters of a typical MLC includi...
متن کاملMonte Carlo modeling and simulations of the High Definition (HD120) micro MLC and validation against measurements for a 6 MV beam.
PURPOSE The most recent Varian(®) micro multileaf collimator (MLC), the High Definition (HD120) MLC, was modeled using the BEAMNRC Monte Carlo code. This model was incorporated into a Varian medical linear accelerator, for a 6 MV beam, in static and dynamic mode. The model was validated by comparing simulated profiles with measurements. METHODS The Varian(®) Trilogy(®) (2300C/D) accelerator m...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Physics in medicine and biology
دوره 55 17 شماره
صفحات -
تاریخ انتشار 2010