Comparison and prediction of pullout strength of conical and cylindrical pedicle screws within synthetic bone

نویسندگان

  • Wen-Chi Tsai
  • Po-Quang Chen
  • Tung-Wu Lu
  • Shing-Sheng Wu
  • Kao-Shang Shih
  • Shang-Chih Lin
چکیده

BACKGROUND This study was designed to derive the theoretical formulae to predict the pullout strength of pedicle screws with an inconstant outer and/or inner diameter distribution (conical screws). For the transpedicular fixation, one of the failure modes is the screw loosening from the vertebral bone. Hence, various kinds of pedicle screws have been evaluated to measure the pullout strength using synthetic and cadaveric bone as specimens. In the literature, the Chapman's formula has been widely proposed to predict the pullout strength of screws with constant outer and inner diameters (cylindrical screws). METHODS This study formulated the pullout strength of the conical and cylindrical screws as the functions of material, screw, and surgery factors. The predicted pullout strength of each screw was compared to the experimentally measured data. Synthetic bones were used to standardize the material properties of the specimen and provide observation of the loosening mechanism of the bone/screw construct. RESULTS The predicted data from the new formulae were better correlated with the mean pullout strength of both the cylindrical and conical screws within an average error of 5.0% and R2 = 0.93. On the other hand, the average error and R2 value of the literature formula were as high as -32.3% and -0.26, respectively. CONCLUSION The pullout strength of the pedicle screws was the functions of bone strength, screw design, and pilot hole. The close correlation between the measured and predicted pullout strength validated the value of the new formulae, so as avoid repeating experimental tests.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: A comparative study between cannulated screws with cement injection and solid screws with cement pre-filling

BACKGROUND Pedicle screws with PMMA cement augmentation have been shown to significantly improve the fixation strength in a severely osteoporotic spine. However, the efficacy of screw fixation for different cement augmentation techniques, namely solid screws with retrograde cement pre-filling versus cannulated screws with cement injection through perforation, remains unknown. This study aimed t...

متن کامل

The Comparison of Pullout Strengths of Various Pedicle Screw Designs on Synthetic Foams and Ovine Vertebrae.

AIM One of the most common problems with transpedicular screws is screw pullout. This study was conducted to measure the pullout strengths of newly designed transpedicular screws. MATERIAL AND METHODS The design of the three group screws were conical cored standard pedicle screw (Type A), dual threaded pedicle screw (Type B), dual core and dual threaded pedicle screw (Type C), respectively. P...

متن کامل

A Biomechanical Comparison of Expansive Pedicle Screws for Severe Osteoporosis: The Effects of Screw Design and Cement Augmentation

Expansive pedicle screws significantly improve fixation strength in osteoporotic spines. However, the previous literature does not adequately address the effects of the number of lengthwise slits and the extent of screw expansion on the strength of the bone/screw interface when expansive screws are used with or without cement augmentation. Herein, four designs for expansive pedicle screws with ...

متن کامل

A parametric study of cylindrical pedicle screw design implications on the pullout performance using an experimentally validated finite-element model.

The present study aims to the design of a finite-element model simulating accurately the pullout behaviour of cylindrical pedicle screws and predicting their pullout force. Three commercial pedicle screws, subjected to pure pullout from synthetic bone, were studied experimentally. The results were used for the design, calibration and validation of a finite-element model. Special attention was p...

متن کامل

Effect of Pilot Hole Tapping on Pullout Strength and Insertion Torque of Dual Core Pedicle Screws.

OBJECTIVE To evaluate the influence of pilot hole tapping on pullout resistance and insertion torque of pedicle screws with a conical core. METHODS Mechanical tests using a universal testing machine were performed on pedicle screws with a conical core that were inserted into pedicles in the fifth lumbar vertebra of calves. The insertion torque was measured using a torque meter with a capacity...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • BMC Musculoskeletal Disorders

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2009