Verification of the accuracy of a photon dose‐calculation algorithm
نویسندگان
چکیده
An extensive set of measured data was developed for the purpose of verifying the accuracy of a photon dose-calculation algorithm. Dose distributions from a linear accelerator were measured using an ion chamber in a water phantom and thermoluminescent dosimeters in a heterogeneous anthropomorphic phantom. Test cases included square fields, rectangular fields, fields having different source-to-surface distances, wedged fields, irregular fields, obliquely incident fields, asymmetrically collimated fields with wedges, multileaf collimator-shaped fields, and two heterogeneous density cases. The data set was used to validate the photon dose-calculation algorithm in a commercial radiation treatment planning system. The treatment planning system calculated photon doses to within the American Association of Physicists in Medicine (AAPM) Task Group 53 (TG-53) criteria for 99% of points in the buildup region, 90% of points in the inner region, 88% of points in the outer region, and 93% of points in the penumbra. For the heterogeneous phantoms, calculations agreed with actual measurements to within +/-3%. The monitor unit tests revealed that the 18-MV open square fields, oblique incidence, oblique incidence with wedge, and mantle field test cases did not meet the TG-53 criteria but were within +/-2.5% of measurements. It was concluded that (i) the photon dose calculation algorithm used by the treatment planning system did not meet the TG-53 criteria 100% of the time; (ii) some of the TG-53 criteria may need to be modified, and (iii) the generally stated goal of accuracy in dose delivery of within 5% cannot be met in all situations using this beam model in the treatment planning system.
منابع مشابه
Evaluation of Dose Calculation Algorithms Accuracy for Eclipse, PCRT3D, and Monaco Treatment Planning Systems Using IAEA TPS commissioning tests in a Heterogeneous Phantom
Introduction: The accuracy of dose calculation algorithm (DCA) is highly considered in the radiotherapy sequences. This study aims at assessing the accuracy of five dose calculation algorithms in tissue inhomogeneity corrections, based on the International Atomic Energy Agency TEC-DOC 1583. Material and Methods: A heterogeneous phantom was sc...
متن کاملEvaluation of AAA and XVMC Algorithms for Dose Calculation in Lung Equivalent Heterogeneity in Photon Fields: A Comparison of Calculated Results with Measurements
Aims: The aims of the present work are (1) to evaluate dose calculation accuracy of two commonly used algorithms for 15 MV small photon fields in a medium encompassing heterogeneity and (2) to compare them with measured results obtained from gafchromic film EBT2.Materials and Methods: Authors employed kailwood (Pinus Wallichiana) to mimic lung. Briefly, seven Kailwood plates, each measuring 25x...
متن کاملDose Calculations for Lung Inhomogeneity in High-Energy Photon Beams and Small Beamlets: A Comparison between XiO and TiGRT Treatment Planning Systems and MCNPX Monte Carlo Code
Introduction Radiotherapy with small fields is used widely in newly developed techniques. Additionally, dose calculation accuracy of treatment planning systems in small fields plays a crucial role in treatment outcome. In the present study, dose calculation accuracy of two commercial treatment planning systems was evaluated against Monte Carlo method. Materials and Methods Siemens Once or linea...
متن کاملDosimetric Evaluation of Dose calculation algorithms of Monaco Treatment Planning System in the heterogeneities area
Introduction: In radiation therapy, the accuracy of dose calculations by a treatment planning system (TPS) is important to achieve tumor control and to spare normal tissue. Treatment planning system calculations in the heterogeneous situation may present significant inaccuracies. In this study, three different dose calculation algorithms, pencil beam (PB), collapsed cone (CC), ...
متن کاملEvaluation of CorePlanTM treatment planning system for 6 and 15 MV photon irradiation
Introduction: Radiotherapy is an effective modality to treat esophageal carcinoma. Due to various inhomogenous tissues with a variety of physical and radiological properties which can cause uncertainty in accurate dose calculations in radiation therapy, accuracy of TPS is very important. The aim of this study was to quantify dose calculation accuracy of CoreplanTM TPS in an inh...
متن کاملEvaluation of Dose Calculation Accuracy of Isogray Treatment Planning System in Craniospinal Radiotherapy
Introduction: Craniospinal radiotherapy is a therapeutic technique for central nervous system (CNS) tumors, which requires meticulous attention to technique and dosimetry.Treatment planning system (TPS) is one of the main equipment in radiotherapy; therefore, the evaluation of its accuracy is essential for dose calculation. The present study evaluates the validity of Isogray TPS in craniospinal...
متن کامل