Comparative accuracy of apical biplane cross-sectional echocardiography and gated equilibrium radionuclide angiography for estimating left ventricular size and performance.
نویسندگان
چکیده
To compare measurements of left ventricular size and performance obtained by apical biplane cross-sectional echocardiography (CSE) and gated equilibrium radionuclide angiography (RNA), we studied 70 patients, all of whom had single-plane and 30 of whom had biplane left ventricular cineangiography. Wideangle, phased-array CSE images were obtained from two orthogonal apical views and left ventricular volumes were calculated using a microprocessor-controlled, video light-pen system programmed for a Simpson's rule algorithm. The average CSE end-diastolic volume (EDV) for all 70 patients of 158 ± 56 ml (± SD) was less than the single-plane angiographic value of 176 ± 68 ml (p < 0.001, r = 0.80). The average CSE end-systolic volume (ESV) of 78 ± 57 ml was not different from the angiographic value of 84 ± 70 ml (r = 0.88). The average CSE and single-plane angiographic ejection fraction (EF) values of 55 ± 16% and 57 ± 19% were not different (r = 0.90). In the 30 patients who underwent biplane angiography, the average CSE EDV of 166 + 62 ml was less than the biplane angiographic EDV of 217 ± 87 ml (p < 0.001, r = 0.81), and the average CSE ESV of 89 ± 69 ml was also less than the angiographic value of 114 ± 89 ml (p < 0.001,
منابع مشابه
Gender differences in clinical trials in coronary heart disease: response to drug therapy.
and volumes by real-time, two-dimensional echocardiography. A comparison of cineangiographic and radionuclide techniques. Comparative accuracy of apical biplane cross-sectional echocardiography and gated radionuclide angiography for estimating left ventricular size and performance. Echocardiographic assessment of systolic and diastolic left ventricular function using an automatic boundary detec...
متن کاملSERVO METHOD FOR BP DETERMINATION/Aaslid and Brubakk
left ventricle in dogs. Circulation 60: 1104, 1979 14. Eaton LW, Maughan WL, Weiss JL: Accurate volume determination in the isolated ejecting canine heart from a limited number of two-dimensional echocardiographic cross-sections. (abstr) Am J Cardiol 45: 470, 1980 15. Helak J, Reichek N, Pearlman E, Weber K, Pietra G, Kastor JA: Quantitation of human left ventricular mass and volume by cross-se...
متن کاملComparison of left ventricular volumes and ejection fraction assessment by two-dimensional echocardiography compared with gated myocardial spect in patients with ischemic cardiomyopathy.
AIM The aim of this study was to assess accuracy of echocardiography in the measurement of left ventricular ejection fraction (LVEF), enddiastolic volume (EDV) and endsystolic volume (ESV) compared with gated SPECT. METHODS A total of 70 patients with left ventricular systolic dysfunction were enrolled to study. Median of the time interval between gated SPECT and echocardiography was 1 hour. ...
متن کاملLeft ventricular volume calculations using a multiplanar transoesophageal echoprobe; in vitro validation and comparison with biplane angiography.
BACKGROUND Biplane angiographic and transthoracic echocardiographic volume calculations have shown to be sufficiently reliable in symmetric hearts; however, they are unreliable in the presence of aneurysmatic distortions. Multiplane transoesophageal echocardiography offers unobstructed cross-sectional views of the heart from one stable transducer position with the potential of imaging irregular...
متن کاملCross sectional echocardiographic assessment of left ventricular volume and ejection fraction in patients with tetralogy of Fallot. Comparison with biplane angiographic measurements.
To evaluate the usefulness and accuracy of calculating left ventricular volume and ejection fraction from cross sectional echocardiograms in patients with tetralogy of Fallot, 28 patients were studied within 24 hours of cineangiography. Indexed end diastolic and end systolic volumes were calculated from three different paired echocardiographic projections: (a) the two and four chamber views fro...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Circulation
دوره 63 5 شماره
صفحات -
تاریخ انتشار 1981