A calibrated fluorescence polarization assay for assessment of fetal lung maturity.
نویسنده
چکیده
The mechanism of a new fluorescence polarization method for measuring phospholipid in amniotic fluid is described. The fluorescence polarization of a surfactant fluorescent dye solubilized in amniotic fluid correlates significantly with the L/S ratio. Model studies indicate that this results primarily from partitioning of the dye between endogenous albumin, which causes a high fluorescence polarization, and dispersed phospholipid, which causes a much lower polarization. The fluorescence polarization can be compared with a calibration curve to give the ratio of phospholipid to albumin in the sample. This procedure may prove useful in antenatal assessment of fetal lung maturity.
منابع مشابه
Prospective clinical evaluation of an improved fluorescence polarization assay for predicting fetal lung maturity.
We performed a prospective clinical evaluation of our newly developed fluorescence polarization procedure to predict fetal lung maturity (Clin Chem 1986;32:248-54). Net fluorescence polarization was measured at 34 degrees C after a 6.5-min incubation of amniotic fluid with fluorophore. For the 26 cases of neonatal respiratory distress syndrome encountered in 196 deliveries, the net polarization...
متن کاملDegree of unsaturation of the fatty acid chains of phospholipids influences the fluorescence polarization: implications for evaluating fetal lung maturity.
To study the effect of fatty acid chain saturation on the fluorescence polarization assay as a measure of fetal lung maturity, we used purified phospholipids isolated from human amniotic fluid and various commercial phospholipids. We found that the fluorescence polarization value decreased as the concentration of unsaturated fatty acids increased. In contrast, the lecithin/sphingomyelin ratio i...
متن کاملImproved fluorescence polarization assay for use in evaluating fetal lung maturity. I. Development of the assay procedure.
We describe a fluorescence polarization assay for use in predicting fetal lung maturity, which is suitable for the TDx Analyzer (an automated fluorescence polarimeter). The assay requires 0.5 mL of amniotic fluid and approximately 30 min, and involves the fluorophore 1-palmitoyl-2(6-[(7-nitro-2, 1, 3-benzoxadiazol-4-yl)amino]caproyl) phosphatidylcholine. Reproducible polarization measurements d...
متن کاملImproved fluorescence polarization assay for use in evaluating fetal lung maturity. II. Analytical evaluation and comparison with the lecithin/sphingomyelin ratio.
We assessed the analytical performance of an improved fluorescence polarization assay for use in evaluating fetal lung maturity and compared results with the lecithin/sphingomyelin ratio. During a three-month period 150 patients' samples were assayed by clinical laboratory personnel with both techniques. Values for the lecithin/sphingomyelin ratio correlate closely with net fluorescence polariz...
متن کاملFetal lung maturity assessed by fluorescence polarization: evaluation of predictive value, correction for endogenous fluorescence, and comparison with L/S ratio.
The steady-state polarization (or anisotropy) of the fluorescent dye 1,6-diphenyl-1,3-5-hexatriene in amniotic fluid samples and the lecithin/sphingomyelin ratio of the samples were correlated with development of respiratory distress syndrome in newborns. We found that clinical samples have a variable endogenous fluorescence that reduces the observed polarization (or anisotropy). This backgroun...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 33 7 شماره
صفحات -
تاریخ انتشار 1987