Estimation of patient setup uncertainty using BrainLAB Exatrac X‐Ray 6D system in image‐guided radiotherapy
نویسندگان
چکیده
The purpose of this study was to evaluate setup uncertainties for brain sites with ExacTrac X-Ray 6D system and to provide optimal margin guidelines. Fifteen patients with brain tumor were included in this study. Two X-ray images with ExacTrac X-Ray 6D system were used to verify patient position and tumor target localization before each treatment. The 6D fusion software first generates various sets of DRRs with position variations in both three translational and three rotational directions (six degrees of freedom) for the CT images. Setup variations (translation and rotation) after correction were recorded and corrected before treatment. The 3D deviations are expressed as mean ± standard deviation. The random error (Σ(σi)), systematic error (μi), and group systematic error (M(μi)) for the different X-ray were calculated using the definitions of van Herk.(1) Mean setup errors were calculated from X-ray images acquired after all fractions. There is moderate patient-to-patient variation in the vertical direction and small variations in systematic errors and magnitudes of random errors are smaller. The global systematic errors were measured to be less than 2.0 mm in each direction. Random component of all patients are smaller ranging from 0.1-0.3 mm small. The safety margin (SM) to the lateral, is 0.5 mm and 2.6 mm for van Herk(1) and Stroom et al.,(2) respectively, craniocaudal axis is 1.5 mm and 3.4 mm, respectively, and with respect to the antero-posterior axis, 2.3 mm and 3.9 mm. Daily X-ray imaging is essential to compare and assess the accuracy of treatment delivery to different anatomical locations.
منابع مشابه
Analysis of the Setup Uncertainty and Margin of the Daily ExacTrac 6D Image Guide System for Patients with Brain Tumors
This study evaluated the setup uncertainties for brain sites when using BrainLAB's ExacTrac X-ray 6D system for daily pretreatment to determine the optimal planning target volume (PTV) margin. Between August 2012 and April 2015, 28 patients with brain tumors were treated by daily image-guided radiotherapy using the BrainLAB ExacTrac 6D image guidance system of the Novalis-Tx linear accelerator....
متن کاملQuantitative evaluation of patient setup uncertainty of stereotactic radiotherapy with the frameless 6D ExacTrac system using statistical modeling
The purpose of this study is to evaluate patient setup accuracy and quantify indi-vidual and cumulative positioning uncertainties associated with different hardware and software components of the stereotactic radiotherapy (SRS/SRT) with the frameless 6D ExacTrac system. A statistical model is used to evaluate positioning uncertainties of the different components of SRS/SRT treatment with the Br...
متن کاملStudy of ExacTrac X‐ray 6D IGRT setup uncertainty for marker‐based prostate IMRT treatment
Novalis Tx ExacTrac X-ray system has the 6D adjustment ability for patient setup. Limited studies exist about the setup uncertainty with ExacTrac X-ray system for IMRT prostate treatment with fiducial markers implanted. The purpose of this study is to investigate the marker-based prostate IMRT treatment setup uncertainty using ExacTrac 6D IGRT ability for patient setup. Forty-three patients wit...
متن کاملA novel compound 6D‐offset simulating phantom and quality assurance program for stereotactic image‐guided radiation therapy system
A comprehensive quality assurance (QA) device cum program was developed for the commissioning and routine testing of the 6D IGRT systems. In this article, both the new QA system and the BrainLAB IGRT system which was added onto a Varian Clinac were evaluated. A novel compound 6D-offset simulating phantom was designed and fabricated in the Prince of Wales Hospital (PWH), Hong Kong. The QA progra...
متن کاملLocal control and overall survival after frameless radiosurgery: A single center experience
Introduction Stereotactic radiosurgery (SRS) has been increasingly advocated for 1-3 small brain metastases. The goal of this study was to evaluate the clinical results in patients with brain metastases treated with LINAC-based SRS using a thermoplastic mask (non-invasive fixation system) and Image-Guided Radiotherapy (IGRT). Material and Methods In this single-institution study 48 patients w...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 16 شماره
صفحات -
تاریخ انتشار 2015