Identification and mitigation of stray laser light in the Thomson scattering system on the Madison Symmetric Torus (MST).
نویسندگان
چکیده
The Thomson scattering diagnostic on the Madison Symmetric Torus (MST) records excessive levels of stray Nd:YAG laser light. Stray light saturates the 1064 nm spectral channel in all polychromators, which prevents absolute electron density measurements via Rayleigh scattering calibration. Furthermore, stray light contaminates adjacent spectral channels for r/a ≥ 0.75, which renders the diagnostic unable to make electron temperature measurements at these radii. In situ measurements of stray light levels during a vacuum vessel vent are used to identify stray light sources and strategies for reduction of stray light levels. Numerical modeling using Zemax OpticStudio supports these measurements. The model of the vacuum vessel and diagnostic includes synthetic collection optics to enable direct comparison of measured and simulated stray light levels. Modeling produces qualitatively similar stray light distributions to MST measurements, and quantifies the mitigation effects of stray light mitigation strategies prior to implementation.
منابع مشابه
Large area avalanche photodiode detector array upgrade for a ruby-laser Thomson scattering system
A low-cost upgrade has been implemented on the Madison Symmetric Torus ~MST! ruby-laser Thomson scattering ~TS! system to increase spectral coverage and substantially improve the signal-to-noise ratio ~SNR!. The spectral resolution has been increased from five channels on the long-wavelength side to an 11-channel system, which covers both sides of the laser line. Coverage on both sides of the s...
متن کاملOperation and beam profiling of an up to 200 kHz pulse-burst laser for Thomson scattering.
A new, high-repetition rate laser is in development for use on the Thomson scattering diagnostic on the Madison Symmetric Torus. The laser has been tested at a rate of 200 kHz in a pulse-burst operation, producing bursts of 5 pulses above 1.5 J each, while capable of bursts of 17 pulses at 100 kHz. A master oscillator-power amplifier architecture is used with a Nd:YVO4 oscillator, four Nd:YAG a...
متن کاملAn integrated data analysis tool for improving measurements on the MST RFP.
Many plasma diagnostics contain complementary information. For example, the double-foil soft x-ray system (SXR) and the Thomson Scattering diagnostic (TS) on the Madison Symmetric Torus both measure electron temperature. The complementary information from these diagnostics can be combined using a systematic method based on integrated data analysis techniques, leading to more accurate and sensit...
متن کاملExperimental evidence for a reduction in electron thermal diffusion due to trapped particles.
New high time resolution measurements of the electron thermal diffusion χ(e) throughout the sawtooth cycle of the Madison Symmetric Torus reversed-field pinch have been made by utilizing the enhanced capabilities of the upgraded multipoint, multipulse Thomson scattering system. These measurements are compared to the χ(e) due to magnetic diffusion predicted by using information from a new high s...
متن کاملImprovements to the calibration of the MST Thomson scattering diagnostic.
Calibration of the Madison Symmetric Torus Thomson scattering system has been refined to improve temperature fluctuation measurements. Multiple avalanche photodiodes have been directly calibrated for use as reference detectors during calibration, improving accuracy and ease of use. From the absolute calibration we calculate corrections to the gain for variation in detector operating temperature...